Table 4
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the m246
The m246 is intended for use in an electromagnetic environment in which radiated RF di
-
sturbances are controlled. The customer or the user of the m246 can help prevent electro
-
magnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the m246 as recommended below, according
to the maximum output power of the communications equipment.
Rated
maximum
output
power of
transmitter
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
P
d
2
.
1
=
P
d
3
.
2
=
P
d
2
.
1
=
Summary of Contents for m246