15
2. Treatment
Preparation
The pressure reducer supplied is to be screwed on
manually to the bottle valve with the help of the
corrugated swivel nut in clockwise rotations.
Strictly do not use any tools like pliers, wrench
or others.
The eclosed spiral hose is to be connected to the
pressure reducer with the help of the swivel nut.
By fast connection the spiral hose is to be connec-
ted to the connection marked on the Precise
3000.
The nasal canula has to be fixed to the connec-
tion marked .
After this the nasal canula is fitted convenient-
ly. To do this, insert the nasal olives into the no-
strils and fix the hose behind the ears with both
hands.
3. Operation and
Settings
3.1. Switching On
If the apparatus was stored at room temperatures
below 10°C the device needs to adapt to the room
temperature, otherwise malfunctions might turn
up.
Open the valve of the bottle slowly. Switch on
the prepared
Precise 3000
by means of the
switch
I/O
. There will be a short operational
check. The previous settings are indicated and do
apply.
The
Precise 3000
can only work correctly if
the user inhales through the nose.
3.2. Power Steps
By pressing the buttons marked
< >
the perfor-
mance steps can be changed in 0.5 l/min steps and
perfectly adjusted to the corresponding situation
in this way.
The performance steps can be changed during
use. See table 1
3.3. Select Operation Mode
Two different operation modes can be set on the
Precise 3000. To do this press the two buttons
marked
< >
simultaneously and keep them pres-
sed. After about 3 seconds the indication in the
display changes from
MANUAL
to
AUTOMA-
TIC
or from
AUTOMATIC
to
MANUAL
.
3.3.1. Operation Mode MANUAL
When set to manual operation mode the amount
of oxygen set with the help of the buttons marked
< >
and indicated in the display is led to the user.
The value’s setting is fixed and remains, inde-
pendent from the users’s perfomance levels.
3.3.2. Operation Mode AUTOMATIC
When the breathing frequency is smaller or exac-
tly fifteen breaths per minute the 02 amount set
with the buttons marked
< >
and indicated in the
display is led to the user.
✗
When the breathing
frequency is above 15 and below 30 breaths per
minute, the oxygen is increased in a linear way.
●
This only happens up the double amount of O
2
set with the buttons marked
< >
not surpassing
the maximum amount of 8 l/min!
✗
The O
2
amount is indicated in the display by
full blocks:
0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 l/min
currently supplied amount = 2 l/min
●
The additional amount supplied is indicated
by half blocks:
0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 l/min
currently supplied amount = 3 l/min
3.4. Responsiveness
The
Precise 3000
automatically adjusts the
sensitivity of the trigger individuel to the user.
Summary of Contents for PRECISE 3000
Page 11: ...Technical Manual and Instructions for Use PRECISE 3000 Electronic gas conserver ...
Page 19: ...Description de l appareil PRECISE 3000 Système économiseur de gaz à contrôle électronique ...
Page 28: ......
Page 29: ...Technische handleidingen Gebruiksinstructies PRECISE 3000 ...
Page 38: ...38 medicare homecare GmbH Hoherodskopf Str 22 D 35327 Ulrichstein ...
Page 39: ......
Page 40: ...40 ...