5640.A
7
Cleaning Instructions
1.
Turn the TEMP 2
off and disconnect it from the power source.
2.
Wipe down the cabinet with a damp cloth using a general hospital cleaner (i.e. Isopropyl
Alcohol) as needed, following the manufacturers instructions for the cleaner.
3.
Avoid contact with the electrical connections and the electrical components.
4.
In the event that excessive fluids enter the cabinet, disconnect power to the unit and refer to
qualified service personnel. Remove the service panel (qualified person) and remove any
liquid and/or dry components. Perform any necessary hospital electrical safety checks before
returning the TEMP 2
to service.
Trouble Shooting
1.
If the unit fails to warm, check the following:
a.
Check to see that the TEMP 2
is plugged in and the power switch is on.
b.
Check the set point setting so that the controller is calling for the unit to operate.
2.
If the unit overheats, call Medical Solutions, Inc.
3.
Contact Medical Solutions, Inc. for technical assistance at 800-435-7600.
Electrical Schematic