EN
Guidance and manufacturer’s declaration – electromagnetic emissions
The Mediblink Compact Nebulizer M440 (JLN-2313AS) is intended for use in the electromagnetic
environment specified below. The customer or the user of the JLN-2313AS should assure that it is
used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The JLN-2313AS uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The The JLN-2313AS is suitable for use in all
establishments including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies
EMC tables
15