88 System Processing Description
z
IBP
unit:
measures
Invasive
blood
pressure
waveform
data.
Figure
42.
IBP
Unit
Block
Diagram
z
Analog
control
unit:
analog
circuit
control,
PIC16C74B
–
selection
of
ECG
channels
and
filter,
size
adjustment,
verification
of
lead
off,
QRS,
heart
rate,
etc.
Isolation DC/DC
Photocoupler
DC +5V
Serial Communication
Isolation DC +/-5V
PIC16F74
ECG Filter Selection
ECG Gain Control
ECG Lead Off
ECG Signal Saturation
Pace Pulse Detection
Respiration Gain Control
Patient Isolation
Figure
43.
Analog
Control
Unit
Block
Diagram
External
Interface
z
TCP/IP
:
Network
connector
z
Serial
I/O
:
a
female
9
pin
sub
‐
miniature
D
shell
connector