40 Service Menu and Factory Default Settings
YM6000 Service Manual
NIBP
Test
These
menus
provide
to
facilitate
performing
verification
testing
for
the
NIBP
subsystem.
For
a
detailed
procedure,
refer
to
Performance
Verification
section
.
Table
12.
NIBP
Test
Tests
Description
Pressure
Sensor
Accuracy
Test
verifies
that
the
pneumatic
pressure
sensor
accuracy
is
within
the
specification.
Air
Leakage
Test
verifies
that
the
pneumatic
pressure
air
leakage
is
within
a
pressure
drop
of
3
mmHg/min.
Inflation
Speed
Test
is
intended
for
factory
use
only.
Deflation
Speed
Test
is
intended
for
factory
use
only.
Return
returns
to
Service
Menu
after
the
test
completed.
Demo
Mode
The
purpose
of
Demo
Mode
is
to
show
a
visual
presentation
demonstrating
how
the
monitor
works.
EtCO
2
Calibration
The
EtCO
2
calibration
may
be
performed
periodically
to
ensure
product
performance
and
measurement
accuracy.
This
menu
provides
to
facilitate
performing
EtCO
2
calibration.
For
a
detailed
procedure,
refer
to
Performance
Verification
section.
Factory
Default
Settings
Factory
default
settings
are
divided
into
three
groups,
adult,
pediatric
and
neonatal
as
described
in
Table
13.
The
patient
mode
is
preset
to
“Adult”
mode.
Alarm
limits
settings
will
be
automatically
changed
to
the
default
settings
for
each
patient
mode
as
the
mode
is
changed
to
Adult,
Pediatric
or
Neonatal
mode.
Table
13.
Factory
Default
Settings
Factory
Defaults
Parameters
Ranges/Selections
Adult
Pediatric
Neonatal
NIBP
NIBP
Initial
Inflation
Pressure
adult/pediatric
120,
140,
160,
180,
200,
220mmHg
(16.0,
18.7,
21.3,
24.0,
26.7,
29.3
kPa)
neonatal
80,
100,
120,
140mmHg
(13.3,
16.0,
17.3,
18.7
kPa)
180
mmHg
(24.0
kPa)
180
mmHg
(24.0
kPa)
120
mmHg
(16.0
kPa)
BP
On
Alarm
On,
Off
Off
Off
Off
NIBP
Interval
Off,
Cont,
1,
2,
2.5,
3,
5,
10,
15,
20,
30,
45,
60,
90,120
,180
minutes
Off
Off
Off
NIBP
SYS
High
Alarm
Limits
adult/pediatric
35
to
270
mmHg
(4.7
to
36.0
kPa)
neonatal
45
to
130
mmHg
(6.0
to
17.3
kPa)
(5
mmHg,
0.7kPa
steps)
160
mmHg
(21.3
kPa)
120
mmHg
(16.0
kPa)
90
mmHg
(12.0
kPa)
NIBP
SYS
Low
Alarm
Limits
adult/pediatric
30
to
265
mmHg
(4.0
to
36.3
kPa)
neonatal
40
to
125
mmHg
(5.3
to
16.7
kPa)
90
mmHg
(12.0
kPa)
70
mmHg
(9.3
kPa)
40
mmHg
(5.3
kPa)
Summary of Contents for YM6000
Page 6: ...iv Contents YM6000 Service Manual This page is intentionally left blank...
Page 14: ...8 Introduction YM6000 Service Manual This page is intentionally left blank...
Page 66: ...60 Troubleshooting YM6000 Service Manual This page is intentionally left blank...
Page 82: ...76 Disassembly Guide YM6000 Service Manual This page is intentionally left blank...