User guide 1533-2001-B
56 of 64
Version 2.0.0 - 2021
Medema A/S
Lamps
Causes
Remedy
6 lamps
The S200 is prevented from driv-
ing. Inhibit 2 is active. This may
be due to the battery charger
being connected, the plug to the
thermal switches being discon
-
nected, a fault in the thermal
switch or the Mini Crosser is
overheating.
The Mini Crosser will turn itself
off automatically after it has
been stationary for 10 minutes.
The battery indicator flashes
with 6 lamps at a rate of about 3
seconds.
Stop the unit and wait 3-5 min
-
utes before restarting. Turn the
key to 0 and then to 1 to reset
the fault.
Contact supplier.
To start X1 again, the key must
be turned to position 0 and back
to I.
7 lamps
Accelerator fault.
Make sure that the accelera-
tor is in neutral when the Mini
Crosser is turned on.
8 lamps
Controller fault.
Contact supplier.
9 lamps
Poor connection to motor brake.
Magnetic brake is disengaged.
Check the connections from
brake to controller, or push the
brake lever up.
Pull the disengagement lever
up for engaging the brakes, for
driving.
10 lamps
The controller has been supplied
with a high voltage in excess of
38 V. Usually seen in the case of
a poor connection.
Check all connections from the
batteries to the controller.
Running
lights
Charger is connected.
Remove charger.
Running
lights from
center.
The throttle has been activated
during start up.
Turn the key to 0 and back to 1
to reset.
Programming
NOTE! For safety reasons, modifications may only be made by
trained personnel, such as Mini Crosser service engineers and
consultants or authorised service personnel at mobility centres,
etc.
Summary of Contents for Mini Crosser X1
Page 1: ...User Manual Mini Crosser X1 1533 2001 B ver 2 0 0 May 2021 GB medemagroup...
Page 61: ...User guide 1533 2001 B 61 of 64 Version 2 0 0 2021 Medema A S Own notes...
Page 62: ...User guide 1533 2001 B 62 of 64 Version 2 0 0 2021 Medema A S...
Page 63: ...User guide 1533 2001 B 63 of 64 Version 2 0 0 2021 Medema A S...
Page 64: ...medemagroup...