User guide P9-0158-B
53 of 60
Version 3.0.3 - 2016
Medema A/S
Status light flashes with 9 flashes per
cycle.
Electronic fault. Contact a skilled
technician. (HMC or therapist).
Status light flashes with 10 flashes per
cycle.
Electronic fault. Contact a skilled
technician. (HMC or therapist).
Status light flashes with 11 flashes per
cycle.
The motor has exceeded the
maximum voltage, or has come close
to exceeding it for longer than the
parameter value. Contact a skilled
technician. (HMC or therapist).
Status light flashes with 12 flashes per
cycle.
Wrong combination of modules.
Contact a skilled technician. (HMC or
therapist).
The Mini Crosser runs at half speed or
not at all.
If the cause is not a raised seat, this
may be a serious fault. Contact a
skilled technician as soon as possible.
(HMC or therapist).
The Dynamic electronic system is programmable so that the
driving characteristics can be optimised for individual users.
Programming must only be carried out by people trained in
Dynamic’s control systems. Changing parameters incorrectly
can produce driving characteristics that are dangerous for the
user.
Note
Mobile phones should not be used very close to the Mini
Crosser as they generate electromagnetic fields that may affect
the electronic system.
The electronic system of the Mini Crosser itself also generates
some electromagnetic fields, which on rare occasions may
affect equipment like the alarm systems in shops.
Dynamic’s control systems have been tested and meet the
requirements of ISO7176/14 and EN12184