Taking the unit out of operation
Disassembly
40
GA 5752 2300 GB 26
5
5.3
Disassembly
5.3.1
Detaching tubes
The disassembly of the tubes is described using the septic fluid jar cap (REF 5750 5362) as an
example.
11
22
Fig. 30: Disassembly of tubes
Switch off the aspirator.
Remove the connection tube (1) from the
black connection of the septic fluid jar cap.
Remove the suction tube (2) from the light
chromed tube connector on the patient
side.
Remove the suction set from the equipment
mount connection.
5.3.2
Removing the overflow protection device
NOTE
Avoid damages to the edge of the float.
11
22
33
Fig. 31: Disassembling the universal overflow pro-
tection
Universal overflow protection
Screw off cap (1) from the overflow
protection device.
Detach float cage (2) and remove float (3).
Summary of Contents for 5752 4856