Introduction
Basic requirements
14
GA 5752 2300 GB 26
1
Sterility:
No sterile product
Single-use product /
reprocessing:
The device and parts of the accessories are reusable. For
information on reprocessing, cleaning and disinfection, please see
the operating instructions.
1.5.3.1
For use in cardiac surgery and surgery of the central nervous system (CNS)
The basic version of the product is a BF protection class device. The unit must be configured to
be CF-proof in order to be used for cardiac or CNS applications.
The following guidelines are therefore imperative:
• A hydrophobic bacterial filter must be fitted between the septic fluid jar and the pump inlet
(REF 5752 1783 or REF 5752 4514).
• All touchable components of the septic fluid jar must be made of non-conductive material
(plastic), for example, cap plug (REF 5752 0184), Serres
®
disposable suction liner 3 litre (REF
310.0401.0 or 310.0411.0) and Serres
®
outer container 3 litre (REF 310.0403 0).
• Only CF-proof suction lances made of a plastic that do not conduct electricity may be used.
• All other equipment parts and accessories must also be CF-proof.
1.5.3.2
BORA variants
These operating instructions apply to the versions listed below:
BORA UP 2080 basic equipment 230 V AC (REF 5752 2296)
• Mains cable 4 m
• Tube adapter
• Filter papers (10 pieces)
• Two caps for covering the mechanical connections
BORA UP 2080 basic equipment 115 V AC (REF 5752 2302)
• Mains cable 4 m
• Tube adapter
• Filter papers (10 pieces)
• Two caps for covering the mechanical connections
BORA UP 2080 OP basic equipment 230 V AC (REF 5752 2301)
• Mains cable 4 m
• Tube adapter
• Filter papers (10 pieces)
• Two caps for covering the mechanical connections
• Pneumatic foot switch
BORA UP 2080 OP basic equipment 115 V AC (REF 5752 2303)
• Mains cable 4 m
• Tube adapter
• Filter papers (10 pieces)
• Two caps for covering the mechanical connections
• Pneumatic foot switch