Initial operation
Overflow protection device
26
GA 5752 2300 GB 26
3
3.3.1.1
Mounting the hydrophobic bacterial and viral filter
NOTE
When mounting the aspirator to an equipment rail, attach a tube with an 8 mm
inner diameter (e.g. REF 5750 5483) using the tube adapter (REF 5752 2295)
between the hydrophobic bacterial and viral filter (1) and the opening (2) on the
back of the aspirator in order to prevent the hydrophobic filter from colliding with
the wall / equipment rail.
11
22
Fig. 9: Inserting the hydrophobic filter
Insert the conical connection (1) of the
hydrophobic filter (1) into the opening (2) on
the back of the device.
Mounting the tubes [
Page 30].
3.3.2
Mounting the surgical overflow protection
11
22
33
55
66
44
Fig. 10: Mounting the surgical overflow protection
Engage float cage (1) with float into lid (2)
of the overflow protection device.
Screw overflow container (3) onto the lid.
Remove cap from the connection of
equipment mount [
Page 29] and push in
equipment mount (4) of the overflow
protection device.
Inserting hydrophobic bacterial and viral
filter (with pore size of 0.2 μm) into
overflow protection device
Remove cap (5) from filter housing
upwards.
Fit the hydrophobic bacterial and viral filter
(6).
Close filter housing with cover.
Mounting tubes [
Page 30].