9.3 Alarming Indicator

High-level 

alarm

High-level 

alarm

Technical 

alarm

Host uses 

Spo2 

sensor

Appendix alarming indicator 

characteristics; Appendix 

alarming sound characteristics;

Display text alarming indication

"Probe shedding".

The screen 

display 

"probe 

shedding" 

Probe 

shedding

Appendix alarming indicator 

characteristics; Appendix 

alarming sound characteristics;

Display text alarming indication

"Probe is short- circuit".

The screen 

display

"Probe is 

short- circuit"

Probe is 

short- circuit

Appendix alarming indicator 

characteristics; Appendix 

alarming sound characteristics;

Display text alarming indication

"low battery".

Battery 

symbol 

remaining 

red

Low battery

Appendix alarming indicator 

characteristics; Appendix 

alarming sound characteristics;

Display text alarming indication

"No pulse".

The screen 

display 

"No pulse”

No pulse

Appendix alarming indicator 

characteristics; Appendix 

alarming sound characteristics;

Display text alarming indication

"Low SpO2".

Measurement 

parameters 

of SpO2 

turn red

Measurement 

parameters 

of PR 

turn red

Measurement 

parameters 

of PR 

turn red

The measurement 
parameters of 
SpO2 are lower 
than the lower 
limit of SpO2 
alarm.

Appendix alarming indicator 

characteristics; Appendix 

alarming sound characteristics;

Display text alarming indication

"High PR".

The measurement 
parameter of PR  
is higher than the 
upper limit of the 
PR alarm

Appendix alarming indicator 

characteristics; Appendix 

alarming sound characteristics;

Display text alarming indication

"Low PR".

The measurement 
parameters of PR 
are lower than the 
lower limit of 
PR alarm

Appendix alarming indicator 

characteristics; Appendix 

alarming sound characteristics;

Display text alarming indication 

"High SpO2".

measurement 

parameters of 

SpO2 

turn yellow

The measurement 
parameters of 
SpO2 are higher 
than the upper 
limit of SpO2 
alarm

Medium-level 

alarm

Alarming Indicator

Quantity

Color

Alarming indicator color

high-level alarm  Red
medium-level alarm  Yellow
low-level alarm    Green

Red

/

Green

2

Attention: 

the alarming level is not changed

Statement: 

the display text alarm indicates that it can be clearly observed at a distance of 1m.

Physiological 

alarm

Connected to the table:

14

This system provides two kinds of inactive alarm shutdown and alarm suspension.
Note

1

The alarm sound continues until the alarm disappears or the alarm is turned off and paused.

2

The alarm signal delay and alarm signal generation delay are less than 1 second.

3

Warning: Do not easily use the alarm shutdown.

9.4 Alarm sound characteristic

9.5 Inactive alarm system

 

Alarming indicator characteristics

Low-level alarm 

Alarm sound feature

Click Alarm pause/

activation Key can 

achieve alarm 

pause

And countdown 

within 120 

seconds

≤120 

seconds

sound 

alarm signal

click Alarm pause/

Activation key Can 

be activated from  

alarm pause or 

waiting for the end 

of pause time

.

Shown as 

chapter 6.6

Shown as 

chapter 6.6

Permanently

closed

Sound alarm 

signal LED 

indication alarm 

signal, 

Text alarm signal

Sound pressure

Alarm sound feature

Sound pressure

Alarm sound feature

Sound pressure

Medium-level alarm

High-level alarm

Low-level alarm 

Medium-level alarm

High-level alarm

Indicator characteristics

Screen tips    Implementation  Pause time   Close content  Activation method

Screen tips    Implementation  Pause time   Close content  Activation method

Occupancy indicator
period
Duty Cycle

Occupancy indicator
period
Duty Cycle
Occupancy indicator
period
Duty Cycle

Green
normally open

100%
Red, Green
About 2 seconds
50%

Red
About 0.5 seconds
50%

The alarm status of the above indicator light can be clearly observed at a distance of 4 m.

Alarm sound pressure measurement: 

the sound pressure meter and the alarm generating device are 

measured at a distance of about 1 m in the same plane.

Alarming Level

Duration
Frequency
Harmonic number
About 55db
Duration
Frequency
Harmonic number
About 60db
Duration
Frequency
Harmonic number
About 62db

0.226 second
250Hz
Not less than 4pcs

About 1.1 second
250Hz
Not less than 4pcs

About 4.38 second
250Hz
Not less than 4pcs

Alarm 

pause

Alarm 

shutdown

Summary of Contents for COX801VB

Page 1: ...eter COX801VB manual for user Please read this manual carefully before using Shenzhen Med link Electronics Tech Co Ltd Address 4th and 5th Floor Building Two Hualian Industrial Zone Xinshi Community D...

Page 2: ...ent process 8 3 Start measuring 8 4 Interpretation of measurement results 8 4 1 SpO2 8 4 2 PR 8 4 3 PI 8 4 4 Bar chart display area 8 4 5 Trace wave display area 8 4 6 Module status display area 8 4 7...

Page 3: ...fied by our company see Appendix P Please note that this product is only used for veterinary It can not be used for human clinical If it is used for human rather than animal by the users and caused th...

Page 4: ...ld for 2 seconds to turn off the power when the power is turned on Figure 1 Type C 4 3 general description of the parameters 4 Working principle COX801VB can continuously measure arterial oxygen satur...

Page 5: ...down key to adjust the value of year then press the set key to switch month day hour minute successively for setting The time shows the 24 hour system 6 1 Time setting 1 Under the setting interface pr...

Page 6: ...t alarming value of PR 1 If the upper limit of the alarming line is not valid i e the lower limit of the alarming line can be arbitrarily selected from 30 bpm to 350 bpm or invalid i e 2 If the upper...

Page 7: ...if you are allergic to silicone and ABS plastics Please dispose of used mainframes accessories and packaging materials in accordance with local laws and regulations Figure 7 Figure 8 Correct connecti...

Page 8: ...her signal processing on the display and transmission of parameters is within 1 second 8 4 Explanation of measurement results Figure 12 12 9 Alarming system 1 alarming signal generated by alarming sys...

Page 9: ...m level alarm Yellow low level alarm Green Red Green 2 Attention the alarming level is not changed Statement the display text alarm indicates that it can be clearly observed at a distance of 1m Physio...

Page 10: ...2 lower limit of alarm PR Upper limit of alarm PR lower limit of alarm Battery is too low no 90 140BPM 50BPM Appeared below about 3 65V Long term transportation storage over 3 months 15 10 specificati...

Page 11: ...hrough TYPEC cable Battery is too low No alarm Alarm pause alarm shutdown or horn failure Please check if the alarm tone is off Please check if the set volume is level 0 Invalid button Internal cause...

Page 12: ...tions or use concentration as low as possible 2 Do not submerge the device in liquid 3 Do not pour liquid onto equipment or accessories 4 Do not allow liquid to enter the casing 5 Do not use abrasive...

Page 13: ...warranty period of our equipment is set according to the electronic product after sales service standards stipulated by national laws The warranty period is one year for mainframe and three months fo...

Page 14: ...nly valid in China one card no replacement 7 The accuracy of this product is medical standard To ensure the accuracy of your use after the warranty period the company promises to continue to provide l...

Reviews: