Appendix S: List of accessories

Appendix R: Symbols and signs

Pulse oximetry sensor

TYPEC

Accessory 

specification

Accessory 

model

Using 

range

Measuring 

part

Tongue clip probe

Type-C Charging Cable

warming

603545123

DS-SMB-001

Animal

Warming:

Please use the specified accessories. Using other accessories may cause damage to animals or monitors.

Disposable accessories can only be used once, and repeated use may result in reduced performance or 

cross-contamination. Do not use the accessory if it or it’s package is found to be damaged.

Operating period environmental requirements, etc.: please check attached manual

Earlobe, tongue

Transport storage 

environment temperature

NON

STERILE

SN

-10

55

5

40

80%

0%

INPUT:
5VDC-1A

Unsterilized Packaging 

Please refer to the 

instruction manual

This product can 

be reused

Working transport 

storage environment 

humidity

This symbol indicates that 
electrical and electronic 
equipment waste cannot be 
disposed of as unsorted municipal 
waste and must be recycled 
separately. Dispose of in 
accordance with the hospital 
waste recycling regulations.

Period of use

No including emulsion

manufacturer

Input: DC 5V1A

Production Date

Serial number

electronic product

Environmental 

protection period

Working temperature

Use and transport s

torage atmospheric 

pressure limits

Accessory 

name

Sound pressure 
Pulse frequency

Number of harmonic components ±25db between 2KHZ and 4KHZ

Pulse duration
Sound pressure
Pulse frequency
Number of harmonic components
Pulse duration

17

Appendix H: Hint sound characteristics

Pulse sound

Touch-tone

Hint sound pressure measurement: 

the sound pressure meter and the alarm generating device are measured 

at a distance of about 1 m in the same plane.

About 50db~62db

1000 Hz

Not less than 1 piece

Not less than 100ms

About 50db~62db

About 938Hz,1.2kHz

No requirement

Not less than 80ms

Appendix P: fault resolution

phenomenon

possible reason

          Solution

Note: 

If the above method still cannot be measured normally, please call the after-sales 

service line 400 058 0755 for consultation.

Unable to boot

No pulse sound and 

button sound

Connect the base or charge the device through 

TYPEC cable

Battery is too low

No alarm

Alarm pause, alarm shutdown or horn failure Please check if the alarm tone is off

Please check if the set volume is level 0

Invalid button

Internal cause

Press multiple times, if it is still invalid, then 

contact customer service

No data after 

Bluetooth connection

Communication code error

Please check if the communication mode of the 

setting interface is Bluetooth communication

No data on the 

serial port

Communication code error

Please check whether the communication mode 

of the setting interface is serial communication

Probe  

unconnected

Probe and device are not connected firmly

Make sure the probe and device are securely 

connected and the probe LED will flash. If the 

probe LED does not flash, use another probe

Finding pulse

Pulse is being searched

If the "Finding Pulse" is displayed continuously 

for more than 60 seconds, measure another 

part quietly

Probe off

Probe off

Please check if the measuring part of 

the probe is off

Volume is 0 or the speaker is faulty

Probe failure

Probe may be damaged

Please replace the probe

Unnormal Pulseless

Pulse undetected or probe may be damaged Please replace the measuring part or 

replace the pulse

Overcurrent

Probe may be damaged

Please replace the pulse

Low perfusion

The probe type is not matched, the 

measurement site is low in perfusion, the 

probe is too tight, the hypothermia causes 

capillary contraction, hypovolemia, 

peripheral vascular disease or anemia, and 

the probe is damaged

Choose the right probe for your animal

Please ensure that blood flow is not restricted 

and the probe should not be too tight

ensure temperature of  animal/measurement 

body is not too low

Replace the measurement site, etc

Blood oxygen values 

don’t match clinical 

estimates

Low perfusion

the measuring position of 

the probe is not suitable

Check if the probe is too tight,

Change measurement position

Unnormal SpO2

Low perfusion index,

Inappropriate probe size,

Inappropriate measurement position

Choose position with higher degree of perfusion,

Recommended blood laboratory test comparison

Choose the right probe

No SpO2 

Probe that does not match the device

Inappropriate probe size

The lighting of the environment is too strong

Choose the right probe for your animal, the 

effect of shield the effect on the probe caused 

by Ambient lighting

Summary of Contents for COX801VB

Page 1: ...eter COX801VB manual for user Please read this manual carefully before using Shenzhen Med link Electronics Tech Co Ltd Address 4th and 5th Floor Building Two Hualian Industrial Zone Xinshi Community D...

Page 2: ...ent process 8 3 Start measuring 8 4 Interpretation of measurement results 8 4 1 SpO2 8 4 2 PR 8 4 3 PI 8 4 4 Bar chart display area 8 4 5 Trace wave display area 8 4 6 Module status display area 8 4 7...

Page 3: ...fied by our company see Appendix P Please note that this product is only used for veterinary It can not be used for human clinical If it is used for human rather than animal by the users and caused th...

Page 4: ...ld for 2 seconds to turn off the power when the power is turned on Figure 1 Type C 4 3 general description of the parameters 4 Working principle COX801VB can continuously measure arterial oxygen satur...

Page 5: ...down key to adjust the value of year then press the set key to switch month day hour minute successively for setting The time shows the 24 hour system 6 1 Time setting 1 Under the setting interface pr...

Page 6: ...t alarming value of PR 1 If the upper limit of the alarming line is not valid i e the lower limit of the alarming line can be arbitrarily selected from 30 bpm to 350 bpm or invalid i e 2 If the upper...

Page 7: ...if you are allergic to silicone and ABS plastics Please dispose of used mainframes accessories and packaging materials in accordance with local laws and regulations Figure 7 Figure 8 Correct connecti...

Page 8: ...her signal processing on the display and transmission of parameters is within 1 second 8 4 Explanation of measurement results Figure 12 12 9 Alarming system 1 alarming signal generated by alarming sys...

Page 9: ...m level alarm Yellow low level alarm Green Red Green 2 Attention the alarming level is not changed Statement the display text alarm indicates that it can be clearly observed at a distance of 1m Physio...

Page 10: ...2 lower limit of alarm PR Upper limit of alarm PR lower limit of alarm Battery is too low no 90 140BPM 50BPM Appeared below about 3 65V Long term transportation storage over 3 months 15 10 specificati...

Page 11: ...hrough TYPEC cable Battery is too low No alarm Alarm pause alarm shutdown or horn failure Please check if the alarm tone is off Please check if the set volume is level 0 Invalid button Internal cause...

Page 12: ...tions or use concentration as low as possible 2 Do not submerge the device in liquid 3 Do not pour liquid onto equipment or accessories 4 Do not allow liquid to enter the casing 5 Do not use abrasive...

Page 13: ...warranty period of our equipment is set according to the electronic product after sales service standards stipulated by national laws The warranty period is one year for mainframe and three months fo...

Page 14: ...nly valid in China one card no replacement 7 The accuracy of this product is medical standard To ensure the accuracy of your use after the warranty period the company promises to continue to provide l...

Reviews: