18
Appendix S: List of accessories
Appendix R: Symbols and signs
Pulse oximetry sensor
TYPEC
Accessory
specification
Accessory
model
Using
range
Measuring
part
Tongue clip probe
Type-C Charging Cable
warming
603545123
DS-SMB-001
Animal
Warming:
Please use the specified accessories. Using other accessories may cause damage to animals or monitors.
Disposable accessories can only be used once, and repeated use may result in reduced performance or
cross-contamination. Do not use the accessory if it or it’s package is found to be damaged.
Operating period environmental requirements, etc.: please check attached manual
Earlobe, tongue
Transport storage
environment temperature
NON
STERILE
SN
-10
℃
55
℃
5
℃
40
℃
80%
0%
INPUT:
5VDC-1A
Unsterilized Packaging
Please refer to the
instruction manual
This product can
be reused
Working transport
storage environment
humidity
This symbol indicates that
electrical and electronic
equipment waste cannot be
disposed of as unsorted municipal
waste and must be recycled
separately. Dispose of in
accordance with the hospital
waste recycling regulations.
Period of use
No including emulsion
manufacturer
Input: DC 5V1A
Production Date
Serial number
electronic product
Environmental
protection period
Working temperature
Use and transport s
torage atmospheric
pressure limits
Accessory
name
Sound pressure
Pulse frequency
Number of harmonic components ±25db between 2KHZ and 4KHZ
Pulse duration
Sound pressure
Pulse frequency
Number of harmonic components
Pulse duration
17
Appendix H: Hint sound characteristics
Pulse sound
Touch-tone
Hint sound pressure measurement:
the sound pressure meter and the alarm generating device are measured
at a distance of about 1 m in the same plane.
About 50db~62db
1000 Hz
Not less than 1 piece
Not less than 100ms
About 50db~62db
About 938Hz,1.2kHz
No requirement
Not less than 80ms
Appendix P: fault resolution
phenomenon
possible reason
Solution
Note:
If the above method still cannot be measured normally, please call the after-sales
service line 400 058 0755 for consultation.
Unable to boot
No pulse sound and
button sound
Connect the base or charge the device through
TYPEC cable
Battery is too low
No alarm
Alarm pause, alarm shutdown or horn failure Please check if the alarm tone is off
Please check if the set volume is level 0
Invalid button
Internal cause
Press multiple times, if it is still invalid, then
contact customer service
No data after
Bluetooth connection
Communication code error
Please check if the communication mode of the
setting interface is Bluetooth communication
No data on the
serial port
Communication code error
Please check whether the communication mode
of the setting interface is serial communication
Probe
unconnected
Probe and device are not connected firmly
Make sure the probe and device are securely
connected and the probe LED will flash. If the
probe LED does not flash, use another probe
Finding pulse
Pulse is being searched
If the "Finding Pulse" is displayed continuously
for more than 60 seconds, measure another
part quietly
Probe off
Probe off
Please check if the measuring part of
the probe is off
Volume is 0 or the speaker is faulty
Probe failure
Probe may be damaged
Please replace the probe
Unnormal Pulseless
Pulse undetected or probe may be damaged Please replace the measuring part or
replace the pulse
Overcurrent
Probe may be damaged
Please replace the pulse
Low perfusion
The probe type is not matched, the
measurement site is low in perfusion, the
probe is too tight, the hypothermia causes
capillary contraction, hypovolemia,
peripheral vascular disease or anemia, and
the probe is damaged
Choose the right probe for your animal
,
Please ensure that blood flow is not restricted
and the probe should not be too tight
ensure temperature of animal/measurement
body is not too low
Replace the measurement site, etc
Blood oxygen values
don’t match clinical
estimates
Low perfusion
,
the measuring position of
the probe is not suitable
Check if the probe is too tight,
Change measurement position
Unnormal SpO2
Low perfusion index,
Inappropriate probe size,
Inappropriate measurement position
Choose position with higher degree of perfusion,
Recommended blood laboratory test comparison
Choose the right probe
No SpO2
Probe that does not match the device
Inappropriate probe size
The lighting of the environment is too strong
Choose the right probe for your animal, the
effect of shield the effect on the probe caused
by Ambient lighting