Med-link Electronics COX801VB User Manual Download Page 7

Warning

9

(1)Select the  appropriate spo2 sensor according to the type, weight and wearing position of the 
tested animal.
(2)Insert the sensor plug into the socket of the monitor.
(3)Clean and disinfect the parts of the sensor in contact with the tested animal.
(4)Connect the sensor to the contact part of the tested animal.
(5)Start measuring.
The Spo2 sensor is connected to the equipment as shown in the following figure. If the 
equipment is not connected to the spo2 sensor or if the spo2 sensor is inserted incorrectly, the 
"probe is not connected" will be displayed at the top of the equipment screen.

8.2 Measurement process

Explosion hazard—Please do not use this product in a flammable or 
explosive gas environments, such as anesthetic or nitrous oxide.
Please do not use this product during animal MRI and CT scans.
Please do not use this product if you are allergic to silicone and ABS plastics.
Please dispose of used mainframes, accessories and packaging materials 
in accordance with local laws and regulations.

Figure 7

Figure 8: Correct connection

Prompt the user to  properly connect the Spo2 sensor:

10

When the signal strength indicates "weak signal ", there is a "?" symbol appearing around the 
SpO2. Indicates that the signal quality obtained by the Spo2 sensor is too poor or too weak. The 
condition of the animal under test should be checked and the sensor placed in a suitable 
location.

Data display interface
After connecting the Spo2 sensor for a few seconds, the measurement result is displayed 
on the screen.
Horizontal vertical device can switch the display direction of the screen
The host uses a pulse carbon oxygen oximeter sensor to display content, range, resolution, 
accuracy, and signal strength. See Appendix B.

8.3 Start measuring

Figure 9

Figure 10

Figure 11

If the tested part and the probe are not in full contact,
"probe shedding" is displayed at the top of the screen

weak signal

Summary of Contents for COX801VB

Page 1: ...eter COX801VB manual for user Please read this manual carefully before using Shenzhen Med link Electronics Tech Co Ltd Address 4th and 5th Floor Building Two Hualian Industrial Zone Xinshi Community D...

Page 2: ...ent process 8 3 Start measuring 8 4 Interpretation of measurement results 8 4 1 SpO2 8 4 2 PR 8 4 3 PI 8 4 4 Bar chart display area 8 4 5 Trace wave display area 8 4 6 Module status display area 8 4 7...

Page 3: ...fied by our company see Appendix P Please note that this product is only used for veterinary It can not be used for human clinical If it is used for human rather than animal by the users and caused th...

Page 4: ...ld for 2 seconds to turn off the power when the power is turned on Figure 1 Type C 4 3 general description of the parameters 4 Working principle COX801VB can continuously measure arterial oxygen satur...

Page 5: ...down key to adjust the value of year then press the set key to switch month day hour minute successively for setting The time shows the 24 hour system 6 1 Time setting 1 Under the setting interface pr...

Page 6: ...t alarming value of PR 1 If the upper limit of the alarming line is not valid i e the lower limit of the alarming line can be arbitrarily selected from 30 bpm to 350 bpm or invalid i e 2 If the upper...

Page 7: ...if you are allergic to silicone and ABS plastics Please dispose of used mainframes accessories and packaging materials in accordance with local laws and regulations Figure 7 Figure 8 Correct connecti...

Page 8: ...her signal processing on the display and transmission of parameters is within 1 second 8 4 Explanation of measurement results Figure 12 12 9 Alarming system 1 alarming signal generated by alarming sys...

Page 9: ...m level alarm Yellow low level alarm Green Red Green 2 Attention the alarming level is not changed Statement the display text alarm indicates that it can be clearly observed at a distance of 1m Physio...

Page 10: ...2 lower limit of alarm PR Upper limit of alarm PR lower limit of alarm Battery is too low no 90 140BPM 50BPM Appeared below about 3 65V Long term transportation storage over 3 months 15 10 specificati...

Page 11: ...hrough TYPEC cable Battery is too low No alarm Alarm pause alarm shutdown or horn failure Please check if the alarm tone is off Please check if the set volume is level 0 Invalid button Internal cause...

Page 12: ...tions or use concentration as low as possible 2 Do not submerge the device in liquid 3 Do not pour liquid onto equipment or accessories 4 Do not allow liquid to enter the casing 5 Do not use abrasive...

Page 13: ...warranty period of our equipment is set according to the electronic product after sales service standards stipulated by national laws The warranty period is one year for mainframe and three months fo...

Page 14: ...nly valid in China one card no replacement 7 The accuracy of this product is medical standard To ensure the accuracy of your use after the warranty period the company promises to continue to provide l...

Reviews: