Summary of Contents for CONCERTO Compressed

Page 1: ...te GmbH Worldwide Headquarters F rstenweg 77a A 6020 Innsbruck Austria Tel 43 512 28 88 89 Fax 43 512 29 33 81 office medel com www medel com CONCERTO Coch le ar I m pl a nt Standard Medium Compresse...

Page 2: ...AW 7846 Rev 1 0 2...

Page 3: ...AW 7846 Rev 1 0 3 1 2 3 4 5 6 7 EN Opening instruction CS Instrukce k otev en...

Page 4: ...0 17 3 CONCERTO Compressed Compressed Electrode array S 15 0 5 1 1 12 1 0 8 Compressed electrode array 4 5 0 9 3 3 15 Medium Electrode 45 7 126 25 4 29 0 17 3 CONCERTO Medium 4 5 0 9 3 3 15 Medium El...

Page 5: ...EXEAS Electrode 114 45 7 25 4 29 0 17 3 CONCERTO FLEXEAS FLEXEAS Electrode array F 24 20 9 1 9 0 8 0 5 0 3 FLEXEAS electrode array Typical dimensions in mm Recommended diameter of cochleostomy Recomme...

Page 6: ...STRO application software user manual SPECIFICATION AND CHARACTERISTICS FOR EACH LEAD AND ELECTRODE ARRAY The implant has 24 independent current sources stimulating 12 independent electrode channels i...

Page 7: ...cause of deafness is non functionality of the auditory nerve and or the upper auditory pathway if external or middle ear infections are present or if the tympanic membrane is perforated in the ear to...

Page 8: ...he round window and its membrane remain intact during drilling the cochleostomy The recommended diameter of the cochleostomy is 1 3 mm for the Cochlear Implant CONCERTO Standard and 0 8 mm for the Coc...

Page 9: ...un sequences in the first level operating mode Before performing an MRI examination all external parts of the CI system must be removed MRI is not completely without risk to the implant and the patien...

Page 10: ...t mu lze ov it proveden m telemetrie viz u ivatelskou p ru ku softwaru aplikace MAESTRO SPECIFIKACE A CHARAKTERISTIKY JEDNOTLIV CH SOUSTAV SVOD A ELEKTROD Implant t m 24 nez visl ch zdroj proudu stimu...

Page 11: ...m vyvinutu kochleu pokud je p inou hluchoty nefunk nost sluchov ho nervu nebo horn sluchov dr hy pokud je v uchu ve kter m m b t provedena implantace p tomna infekce zevn ho nebo st edn ho ucha nebo p...

Page 12: ...nimalizov no riziko poopera n infekce je nutno db t na to aby kulat ok nko a membr na z staly b hem vrt n kochleostomie intaktn Doporu en pr m r kochleostomie je 1 3 mm pro kochle rn implant t CONCERT...

Page 13: ...rovozn m re imu prvn ho stupn P ed prov d n m vy et en MRI je nutno odstranit v echny extern sti kochle rn ho implant tu MRI nen pro implant t a pro pacienta pln bez rizika V p pad pou it MRI je nutno...

Page 14: ...AW 7846 Rev 1 0 14...