PHM-104A HIGH/LOW SPEED PRECISION PUMP 

 

Table 3.1 - PHM-104A Control Software User Interface 

Pump ID 

Pulldown menu that contains PUMP IDs for each pump that was connected to the computer before the 
application was opened.  Select the desired pump to view/change infusion settings. 

Reset 

Sets the Syringe, Dose, and Rate values to the pump’s default values and causes the PHM-104A Control 
Application to recognize any pumps that have been connected to the computer after the application 
was opened.  Selecting Reset disables the Download and Set Default buttons.  Changing any of the 
values in the Syringe, Dose, and Rate fields will enable these buttons. 

Home 

Returns the pusher block to the home position.  This position varies based on the Syringe, Dose, and 
Rate settings. 

Prime 

Starts an injection from the current position of the pusher block to the end of the syringe then returns 
the pusher block to the Home position. 

Syringe (

µ

L) 

Select the size of the Hamilton MICROLITER™  syringe that is loaded in the PHM-104A.  Possible 
selections are 10, 100 and 500 

µ

L. 

Dose (

µ

L) 

Enter the desired dose (

µ

L) in this field.   

Rate (

µ

L/sec) 

Enter the desired rates (

µ

L/sec) in this field.  See Figure 2.6 dose capabilities. 

OK 

Indicates that the Dose and Rate settings are within acceptable limits.  

Invalid Data 

Indicates that the value entered in the Dose and Rate fields are not within acceptable limits.  Refer to 
Dose Capabilities for these limits.  The value will be reverted back to the last good value used.  

Download 

Click to send Syringe, Dose, and Rate settings to the currently selected pump.  If all settings are valid, 
the Inject and Set Default buttons will be enabled. 

Inject 

Click to inject the dose. 

Set Default 

Enter the desired default values for the pump in the Syringe, Dose and Rate fields and click Download 
to enable the Set Default button.  Click Set Default to save these values as this pump’s defaults.  The 
default values are used when Inject button or the MED-PC input is used after the USB is connected.  

 

7.

 

Select the pump to configure from the 

Pump ID 

list.  If a pump that is connected to the computer 

does not appear on this list, verify that it is properly connected and turned on and click the 

Reset

 

button. 

8.

 

This pump’s default values for Syringe, Dose and Rate will be displayed.  Click 

Inject 

to begin an 

infusion using the default values or change any of these values to enable the 

Download

 and 

Set 

Default

 buttons. 

9.

 

If the Syringe, Dose or Rate values are changed, click 

Download

 to send the new settings to the 

pump.  If these values are within acceptable limits, the 

Inject

  and 

Set Default

  buttons will be 

enabled.  If one or more of these values is not within acceptable limits the Status of the pump will 
indicated as 

Invalid Data

 and the

 Inject

 and 

Set Default

 buttons will not be enabled. 

10.

 

Click 

Inject

 to begin the infusion or 

Set Default

 to save these values as the default settings for this 

pump. 

NOTE:  The Inject and Set Default buttons will be disabled until valid values are entered in the Dose 
and Rate fields and Download is clicked.   

 

 

 

- 9 -

 

DOC-187 Rev. 1.6 Copyright © 2019 

Med Associates, Inc. 

Summary of Contents for PHM-104A

Page 1: ...or research HIGH LOW SPEED PRECISION PUMP PHM 104A USER S MANUAL DOC 187 Rev 1 6 Copyright 2019 All Rights Reserved Med Associates Inc P O Box 319 St Albans Vermont 05478 Phone 802 527 2343 Fax 802 52...

Page 2: ...This page intentionally left blank...

Page 3: ...ose Capabilities 4 Operating Instructions 6 Chapter 3 Syringe Placement 6 Hardware Setup 7 Software Controlled Operation 8 MED PC Controlled Operation 10 USB Control Via MED PC 10 Sample MED PC Progra...

Page 4: ...______________ _____________________________________________________________________________________ _____________________________________________________________________________________ _____________...

Page 5: ...quality Hamilton 1700 Series gastight glass syringes to ensure precise accurate delivery The pump supports three syringe sizes 10 l 100 l and 500 l A wide range of injection rates is possible as well...

Page 6: ...position Enables the MED OPERATE connector When the switch is in TEST position Completes one infusion at the default pump settings refer to Table 3 1 PHM 104A Control Software User Interface Set Defau...

Page 7: ...IATES INC PHM 104A HIGH LOW SPEED PRECISION PUMP Figure 2 2 Manifold End of the PHM 104A Figure 2 3 Control End of the PHM 104A Figure 2 4 Syringe Components 3 DOC 187 Rev 1 6 Copyright 2019 Med Assoc...

Page 8: ...on rate included in the highlighted portion of each graph is attainable Using Figure 5 10 uL syringe as an example a dose of 3 uL may be infused at a rate up to 27 uL sec Dose capabilities are based o...

Page 9: ...igure 2 6 PHM 104A Pump Dose Capabilities 100 uL PHM 104A Infusion Capabilities by Dose 100 uL Syringe Figure 2 7 PHM 104A Pump Dose Capabilities 500 uL PHM 104A Infusion Capabilities by Dose 500 uL S...

Page 10: ...h the retainer in the pusher block and align the syringe barrel with the syringe holder being sure that the syringe tabs are aligned with the retainer 4 Press down firmly the syringe should fit snugly...

Page 11: ...the section of tubing in the infusion solution 4 Connect a section of tubing of the required length to the Output port on the manifold shown in Figure 2 2 Manifold End of the PHM 104A using a female...

Page 12: ...Connecting the PHM 104A pump to a new USB port will require the reinstalling of the device driver 3 Set the FLUSH REMOTE TEST switch to REMOTE 4 Set the desired pump I D using the PUMP ID pushbuttons...

Page 13: ...everted back to the last good value used Download Click to send Syringe Dose and Rate settings to the currently selected pump If all settings are valid the Inject and Set Default buttons will be enabl...

Page 14: ...switches shown in Figure 2 1 Top Panel with Syringe Loaded The pump ID corresponds to the Box in MED PC 3 Set the FLUSH REMOTE TEST switch to REMOTE 4 Using the 24 VDC power supply connect the 24 VOLT...

Page 15: ...t values P104A_INJECT Updates values and begins pump injection cycle Comments Sets the infusion parameters for the pump The maximum and minimum Rates are dependent on the Syringe Size and Dose To dete...

Page 16: ...rogram records no data and has no end criteria P104A_Init Initializes the PHM 104A pump Must be used at the beginning of the program This command only needs to be executed once P104A_PumpSetup Sets th...

Page 17: ...Dipper are ordered It will be necessary to change one of these LeftLight 4 RightLight 5 HouseLight 7 Pump 8 A Control Variables with Assigned Aliases as Defined Var_Alias Dose ul A 0 Default 10 ul Va...

Page 18: ...3 Syringe Size A 2 S2 PUMP CONTROL S S 2 S1 Initialize the Pump 0 1 B P104A_Init MG SHOW 6 Init Resp B S2 S2 Wait for a Response Send the Currently selected values to the Pump R LeftLever C P104A_Pum...

Page 19: ...efault parameters using the pump control application described in Figure 3 4 PHM 104A Control Software Screen 4 Connect the MED OPERATE port Figure 2 3 on the pump to any available output on a standar...

Page 20: ...ation again if the PHM 104A is connected to a different USB port at a later time Insert the PHM 104A CD into the CD ROM drive and the screen shown in Figure 4 1 PHM 104A Main Screen will appear Click...

Page 21: ...ctor on the PHM 104A to a standard wall outlet Using the included USB cable connect the USB port on the PHM 104A to any available USB port on the computer The Found New Hardware Wizard screen will app...

Page 22: ...N Please contact Med Associates Inc for information regarding any of our products For Technical questions email support med associates com For Sales questions email sales med associates com Visit our...

Reviews: