MED ASSOCIATES, INC.
PHM-104A HIGH/LOW SPEED PRECISION PUMP
Dose Capabilities
Figure 2.5 - Figure 2.7 display the PHM-104A infusion capabilities for 10 uL, 100 uL and 500 uL syringes,
respectively. Infusion rates are dependent upon the size of the syringe and the dose volume. The
minimum infusion rate for all three syringe sizes equals 10% of the syringe volume. Any infusion rate
included in the highlighted portion of each graph is attainable. Using Figure 5 (10 uL syringe) as an
example, a dose of 3 uL may be infused at a rate up to 27 uL/sec. Dose capabilities are based on saline
infusion. Pump performance may vary depending upon the viscosity of fluid used. Attempting to pump
extremely viscous fluids, especially at high rates, may lead to inaccurate infusion volumes and pump
faults.
Figure 2.5 – PHM-104A Pump Dose Capabilities 10 uL
PHM-104A Infusion Capabilities by Dose
(10 uL Syringe)
- 4 -
DOC-187 Rev. 1.6 Copyright © 2019
Med Associates, Inc.