MED ASSOCIATES INC.
RETRACTABLE LEVER
- 4 -
DOC-050 Rev 1.4 Copyright © 2017
Instructions
1.
Remove Retractable Lever from chamber and disconnect the cables from back of lever.
2.
Using a #1 Philips screwdriver, remove the 4 screws (2 on each side) shown below.
3.
Slide the case backwards, making sure not to pull completely off as there are wires
connecting the circuit board to the lever assembly.
4.
Spray compressed air into lever, below the paddle, also spraying into all gaps and
crevices to loosen any debris and allowing debris to fall out the lever.