MED Associates ENV- 112CM Product Manual Download Page 6

MED ASSOCIATES INC. 

 

RETRACTABLE LEVER 

- 3 -

 

DOC-050 Rev 1.4 Copyright © 2017 

 

CHAPTER 3

 

|  ADJUSTABLE TENSIONER 

The adjustable lever  tension is factory set to 25 grams but may be adjusted from 15 to 100 

grams. Turn the tensioner clockwise to increase lever tension and counter clockwise to decrease 

lever tension. The tension should not be adjusted without consulting Med Associates Support. 

Figure 3.1 – Tensioner  

  

 

CHAPTER 4

 

| MAINTENANCE 

Cleaning 

It is very important and necessary to schedule periodic cleaning of the ENV-112CM Retractable 

Response Lever to ensure reliable operation and response detection. This should be done once a 

week or once a month based on use, to reduce buildup of dander, dust, or debris which can lead 

to unreliable operation and response detection. Humid environments may require more 

frequent cleaning. Med-Associates recommends that sticky substances such as sucrose, peanut 

butter, etc. should not be placed on the paddles of retractable levers as it can cause damage to 

mechanical and electrical components.  

NOTE: Items returned for repair that have not been properly cleaned will be not covered 

under warranty. Please contact Med Associates (see Appendix A | Contact Information) if 

you would like to send your Retractable Response Lever to us for a routine cleaning 

service. We will be happy to send you a quote for this service. 

Cleaning Supplies/Tools  

• #1 Philips Screwdriver  

• Compressed Air  

• Swab and Wipes  

• Quatricide or 10% Ethanol 

Summary of Contents for ENV- 112CM

Page 1: ...for research RETRACTABLE RESPONSE LEVER ENV 112CM PRODUCT MANUAL DOC 050 Rev 1 4 Copyright 2017 All Rights Reserved Med Associates Inc P O Box 319 St Albans Vermont 05478 Phone 802 527 2343 Fax 802 52...

Page 2: ...i DOC 050 Rev 1 4 Copyright 2017 Med Associates Inc notes...

Page 3: ...apter 1 Introduction 1 Specs 1 Chapter 2 Connections 2 Operate Connector 2 Response Connector 2 Optional Lever Position Connector 2 Chapter 3 Adjustable Tensioner 3 Chapter 4 Maintenance 3 Cleaning 3...

Page 4: ...V 112CM permits placement just 2 1cm above the grid floor and can be raised as needed with chamber filler panel inserts This lever is designed to work with ENV 007 ENV 008 and ENV 009 series chambers...

Page 5: ...n Function Wire Color 1 28 Volt Ground Black 2 Extend Retract Lever White 3 28 Volts Red Response Connector Response connects to any available Input on a Standard MED Connection Panel via an included...

Page 6: ...done once a week or once a month based on use to reduce buildup of dander dust or debris which can lead to unreliable operation and response detection Humid environments may require more frequent clea...

Page 7: ...of lever 2 Using a 1 Philips screwdriver remove the 4 screws 2 on each side shown below 3 Slide the case backwards making sure not to pull completely off as there are wires connecting the circuit boar...

Page 8: ...4 again to blow out any remaining debris 7 Slide the case forward and reattach using the 4 screws removed in step 2 Reinstall lever in chamber APPENDIX A CONTACT INFORMATION Please contact Med Associ...