Diagnostic System and Possible Solutions
the device is provided with a diagnostic circuit that allows identifying the main operating issues.
Signalling Type
Description
Solution
4 Beep
the temperature control
circuit is failed.
the device does not emit
any light.
Contact a Mectron service
centre.
3 Beep
the control circuit indicates
that the maximum LED
temperature has been
reached.
the device does not emit
any light.
Place the device back in
its housing and wait about
10 minutes before using it
again.
Mode LED flashing
the control circuit reports an
lED failure.
the device does not emit any
light, or the intensity of the
emitted light is very low.
Contact a Mectron service
centre.
Table 5 –
Diagnostic system and Possible solutions
7.3
How to Prepare the Device and the Accessories
Before Sending them to Repair
In the event that there is a need to send the device, the optical fiber and accessories to an
authorized Mectron service center, we invite our kind customers to respect the good code of
conduct reported in the following:
1.
Clean the device, optic fibre and accessories according to the instructions reported in the
10 of
the Use and Maintenance manual;
2. sterilse the sterilisable parts according to the instructions reported in the
page 10 of
the Use and Maintenance manual:
•
Optic fibre;
• Optical protection.
3. leave the sterilised parts in their sterilization pouches that certify the sterilization
completion;
4.
If the device is still under warranty, attach a copy of the purchase document;
5. Whenever possible, send all the parts in their original packaging or, alternatively, in
proper packages, to avoid any damage during the transportation.
the above requirements (points 1 and 2) are in compliance with the current requirements
concerning the protection of health and safety in the workplace, as per Italian legislative decrees
81/08 and subsequent amendments.
In the event that the customer does not comply with the requirements (points 1 and 2), Mectron
reserves the right to charge him/her for cleaning and sterilization costs or to reject the items for
repair received in unsuitable conditions.