41
TROuBLESHOOTING
NOTE
: If necessary to achieve the
IMMuNITY TEST LEvEL, the distance
between the transmitting antenna and the
PIEZOSURGERY® plus may be reduced to 1
m. The 1 m test distance is permitted by IEC
61000-4-3.
WARNING:
Portable RF communication
equipment (including peripheral devices as
antenna cables and external antennas) must
not be used closer than 30 cm (12 inches) to
any part of the device PIEZOSURGERY® plus,
including the cables specified by the
manufacturer. Otherwise, there may be a
performance degradation of these devices.
8
TROUBLESHOOTING
8.1
Diagnostic Messages and Solutions
The
PIEZOSURGERY® plus
is provided with a diagnostic circuit that detects any operating
anomalies and display them on the touch screen by using the “MED-XXX” notation.
There are three different levels of event notifications:
Blocking
These conditions cause the device
operation halting, preventing the use of
the specific channel or the whole device
in order to protect the patient.
1
Non-blocking
These conditions temporarily inhibit the
use of the specific channel or the whole
device. The Operator MuST confirm
the reading of the message in order to
proceed.
NOTE:
In the event of a non-blocking
condition, after having confirmed
the reading of the message, it will be
necessary to confirm that the necessary
actions have been carried out in order to
proceed.
2