34
Emissions Test
Compliance
Electromagnetic Environment Guidance
RF Emissions
CISPR 11
Class A
The
PIEZOSURGERY® plus is suitable
for use in all establishments other than
domestic, and may be used in domestic
establishments and those directly
connected to the public low voltage power
supply network that supplies buildings
used for domestic purposes, provided the
following warning is heeded.
CAUTION:
This device is intended for
use by healthcare professionals only.
The EMISSIONS characteristics of the
PIEZOSURGERY® plus
make it suitable
for use in industrial areas and hospitals
(CISPR 11 class A). If it is used in a
residential environment (for which CISPR
11 class B is normally required) the
PIEZOSURGERY® plus
might not offer
adequate protection to radio-frequency
communication services. The user might
need to take mitigation measures ,
such as relocating or reorienting the
PIEZOSURGERY® plus.
Harmonic emissions
IEC 61000-3-2
Class A
Emissions of fluctuations
voltage/flicker
IEC 61000-3-3
Compliant