11
INSTALLATION AND CONFIGuRATION
CAUTION
: Before connecting the
handpiece cord to the device, check that the
electrical contacts are perfectly dry. Dry with
compressed air if necessary.
CAUTION
: Do not allow the machine
body or foot pedal to become wet. If liquid
should enter the machine body or foot pedal,
this may cause damage.
CAUTION
: No changes can be made to
this device.
4.2
Device Installation
Equipotential plug: there is an additional
equipotential plug on the back of the
device. The plug complies with DIN
42801. Insert the equipotential cable
connector (not supplied) onto the plug
on the back of the device. The purpose of
the additional equipotential connection
is to reduce the difference in potential
that may occur during operation between
the device body and the conductive
parts of other objects within the medical
environment;
1
Insert the power supply cord into its
connector on the back of the device and
then into the wall socket;
CAUTION:
Position the device in
such a way as to ensure that the plug
is easily accessible as this is considered
a means for disconnecting the power
supply.
2
Connect the foot pedal to the back of
the device in the socket marked with
the symbol , using the foot pedal
cable plug, ensuring it clicks into place.
To disconnect the foot pedal from the
machine body, grasp the connector, crush
the release sleeve and pull backwards;
NOTE:
if the foot pedal ceases
functioning, the button on the back, left
of the device, below the symbol , can
be used to continue treatment.
3