Settings
Jog settings
Within the jog settings menu, you can configure the jog speed limits while in jog mode. Below is a detailed
breakdown of each setting and the options available for each setting.
Setting
Setting
Action
Action
Range
Range
Clockwise
Speed
Configure the jog speed in a clockwise direction
0.1 to
30
rev/min
Anticlockwise
Speed
Configure the jog speed in an anticlockwise direction
0.1 to
30
rev/min
Jog Timeout
Period
Set the timeout (in minutes) that the machine will keep the motor drive engaged, before
the motor drive is disabled. The load applied to the torque cell, must reach at least 25 %
of the cell's capacity before the timeout activation is applied. At the end of the timeout
period, the 'Jog Active' menu screen is automatically switched back to the 'Ready to Test'
menu screen.
(
Example:
Example:
Vortex-
dV
stand fitted with a 6 N.m torque cell, must reach 1.5 N.m in either
direction, before the timeout activation is applied.
Forces below the 25 % limit will not
activate the timeout and the stand will actively hold the load applied
1 to 59
minutes
Clockwise
Torque Limit
Configure the clockwise torque limit while in jog mode
0 to
125%
torque
cell
capacity
Anticlockwise
Torque Limit
Configure the anticlockwise torque limit while in jog mode
0 to
125%
torque
cell
capacity
Tare AFTI
Configure whether or not the AFTI is tared when pressing the tare button in jog mode.
Yes or
No
Units
Within the units menu, you can configure the test stands units for displacement and speed. Torque settings are
configured using the AFTI.
Setting
Setting Units Available
Units Available
Angle rev, deg
Speed rev/min, rev/sec, deg/min, deg/sec
Torque
N.m, N.cm, mN.m, gf.cm, kgf.cm, kgf.m, lbf.ft, lbf.in,
ozf.in
Edit Test
Important!
Important!
For more information relating to ‘Start Direction’ and ‘Move to Start’, please see
Vortex
Vortex
--
dV
dV
Operation Sequence and Move to Start.
Operation Sequence and Move to Start.
Cycle (by Angular Displacement)
In a cyclic test, the spindle moves between two reference angles that are relative to tared zero.
Setting
Setting
Options
Options
Cycle Count
0-9999
Clockwise Speed
0-30 rev/min (0 to 180 deg/s)
Anticlockwise
Speed
0-30 rev/min (0 to 180 deg/s)