CrimpTest-1 kN Calibration Checking Rig
Mecmesin
9
Interpreting the results
The CrimpTest-1 kN has a permissible accuracy of ±0.2% FSD
(±2 N).
Applied load Lower limit Upper limit
Reading
Pass/Fail
0 N
0 N
0 N
100 N
98 N
102 N
200 N
198 N
202 N
300 N
298 N
302 N
400 N
398 N
402 N
500 N
498 N
502 N
600 N
598 N
602 N
800 N
798 N
802 N
1000 N
998 N
1002 N
Ensure masses used are calibrated for your location, and that
calibration is done in an environment of temperature between
15ºC and 30ºC.
If your readings are outside the limits, please contact your
distributor.