Electromagnetic Flowmeter M921
MEATEST
34
Operation manual
Maximum flowrate value reading
RMX?
Response contains maximum Flowrate value.
Example:
If query „RMX?<cr>” is sent, flowmeter returns response in format „100.000<cr>“ (maximum flowrate value
indicated since last reset – command CLRMM).
Date and time of maximum flowrate reading
RXD?
Response contains time & date of maximum flowrate.
Example:
If query „RXD?<cr>” is sent, flowmeter returns response in format „10:06 15/01/2009<cr>“ (time & date of
maximum flowrate value indicated since last reset – command CLRMM).
Minimum flowrate value reading
RMN?
Response contains minimum Flowrate value.
Example:
If query „RMN?<cr>” is sent, flowmeter returns response in format „0.000 “ (minimum flowrate value
indicated since last reset – command CLRMM).
Flow unit resolution
FFR(?)<CPD> { 0 | 1 | 2 | 3 | 4 | 5 }
Following modes can be set:
0
none (0)
1
1 decimal
2
2 decimals
3
3 decimals
4
4 decimals
5
default
M921 confirms execution with string „Ok<cr>”.
Example:
Command „FFR1<cr>” sets mode “1 decimal” for flowrate resolution. If query „FFR?<cr>” is sent,
flowmeter returns response in format „1<cr>”.
Volume unit resolution
FVR(?)<CPD> { 0 | 1 | 2 | 3 | 4 | 5 }
Following modes can be set:
0
none (0)
1
1 decimal
2
2 decimals
3
3 decimals
4
4 decimals
5
default