miniLAB 1008 User's Guide
Calibrating and Testing the Device
4.
Click the
OK
button to begin calibration. The first of two
Update
Input
Connections
dialog opens.
5.
Connect the analog input pins labeled CH0 IN, CH1 IN, CH2 IN, CH3 IN, CH4 IN,
CH5 IN, CH6 IN, and CH7 IN to the
CAL
output (pin 15) and press
OK
. The
second
Update
Input
Connections
dialog opens.
6.
Connect the analog input channels labeled CH0 IN, CH2 IN, CH4 IN, and CH6 IN
to the
CAL
output (pin 15), and connect the analog input channels labeled CH1 IN,
CH3 IN, CH5 IN, and CH7 IN to one or more
GND
terminals, and then click the
OK
button.
When all of the gain and offset corrections are calibrated, the following dialog
opens.
7.
Click the
OK
button to exit the calibration procedure.