CIO-EXP16 User's Guide
Functional Details
Output voltage vs gain at maximum temperature
Type
Min Temp
°
C Max
Temp
°
C
mV at
Max
Gain / Output Volts @ Max Temp
100 200 300 500 800
J
0
750
42.28 4.2 8.5 12.7 21.1 33.8
K
−
200
1250
50.63 5.1 10.1 15.2 25.3 40.5
T
−
200
350
17.82
1.8 3.6 5.3 8.9 14.3
E
−
200
900
68.78 6.9 13.8 20.6 34.4 55.0
S 0
1450
14.97
1.5 3.0 4.5 7.5 12.0
R 0
1450
16.74
1.7 3.3 5.0 8.4 13.4
The table below lists the maximum readable temperatures for each gain and thermocouple type for an A/D
range of ±5V.
Gain vs maximum readable temperatures
Type
Gain / Max. Readable Temps. with A/D Range = ±5 V
10 100 200 300 500 800
J
MAX MAX 457°C 307°C 186°C 117°C
K
MAX 1232
°C
602°C 406°C 246°C 152°C
T
MAX MAX MAX 330°C 214°C 140°C
E
MAX 661°C 350°C 244°C 152°C 98°C
S
MAX MAX MAX MAX 1035°C
697°C
R
MAX MAX MAX 1445°C
961°C 658°C
To read the maximum usable temperature for type K and E thermocouples, set the gain to
X10
or less.
After you change the gain setting, calibrate the CIO-EXP16 using the
Insta
Cal utility.
21