Straingage half bridge
ch 1
ch 2
ch 3
ch 4
+Us
2
13
24
35
+Ud
3
14
25
36
-Us
5
16
27
38
HBx
8
19
30
41
Sw.
1
12
23
34
The active Straingage R3 and R4 are connected to the terUs,
+Ud and -Us.
HBx: The solder bridges “8”, “19”, “30” or “41” must be closed so that the
internal extension resistors R1 and R2 are activated.
Straingage quarter bridge
ch 1
ch 2
ch 3
ch 4
+Ud
3
14
25
36
-Us
5
16
27
38
HBx
8
19
30
41
AUX
6
17
28
39
QB 120
11
22
33
44
QB 350
10
21
32
43
QB 1000
9
20
31
42
Sw.
1
12
23
34
The active Straingage R4 is connected to the terUd, AUX and
-Us in 3-wire technology.
HBx: The solder bridges “8”, “19”, “30” or “41” must be closed so that the
internal extension resistors R1 and R2 are activated.
A solder bridge should be set depending on the Straingage resistance
(120/350/1000 Ohm), e.g. 120 Ohm at channel 1: Solder bridge 11
(vertical); 350 Ohm at channel 2: Solder bridge 21 (horizontal).
ME-Meßsysteme GmbH
Neuendorfstr. 18a
Tel.: +49 3302 89824 60
Mail: [email protected]
22
16761 Hennigsdorf
Fax: +49 3302 89824 69
Web: www.me-systeme.de
+Us
-Us
-Ud
+Ud
R1
R2
R3
R4
+Us
-Us
-Ud
+Ud
R1
R2
R5
R4
AUX
Summary of Contents for GSV-4
Page 2: ......