Protect the monitor from sudden temperature changes that can lead a condensation within the
instrument. To minimize the risk of condensation, avoid moving the system between heated
buildings and outside storage. Once moved inside, let the monitor in the unopened shipping
container to reach the room ambient temperature before unpacking and placing into service.
Before operating the system, wipe down all visible condensation and allow the system to
reach equilibrium at room temperature.
11.2
Operating Environnement
The NIPE Monitor V1 is not designed for use in areas containing flammable gases or vapors.
WARNING
:
Explosion hazard: do not use the NIPE Monitor V1 in a flammable atmosphere
or where concentrations of flammable anesthetics may occur.
The monitor is designed to operate safely in a temperature from 5°C to 35°C to an altitude of
2000 meters. The conditions that exceed these limits may modify its reliability.
The monitor is designed to operate in a humidity range from 15% to 95% (non-condensing).
The monitor works correctly in conditions of sea level and below, and is not affected by
extremes or by altitude modifications to 2000 meters and a pressure from 480hPa to 1066hPa.
11.3
Power Requirements and System Grounding
The NIPE Monitor V1 requires a power source of 100 to 240 VAC, 50-60Hz.
To protect operating personnel and patients, the monitor must be properly grounded.
Accordingly, the monitor is equipped with a hospital grade power supply cord. The power
cord grounds the system to the power line ground when plugged into an appropriate three-
wire socket.
CAUTION:
FOR PROPER GROUNDING, THE POWER SOCKET MUST BE A THREE-
WIRE GROUNDED OUTLET. A HOSPITAL GRADE OUTLET IS REQUIRED. NEVER
ADAPT THE THREE-PRONG PLUG FROM THE MONITOR TO FIT A TWO-SLOT
OUTLET. IF THE OUTLET HAS ONLY TWO SLOTS, MAKE SURE THAT IT IS REPLACED