NIPE Monitor V1 - Continuous monitoring system of comfort
.
/
.
discomfort for newborn infants
MD/PRD/IN16.NIPEV1 v.10 - 12 MAR 2020
15
The other functions available from the main menu (figure 4) are presented in the following
chapters:
-
Maintenance : cf. Chapter 8.2
-
Records management : cf. Chapter 8.3
-
Events update : cf. Chapter 8.4
-
Date and Time settings : cf. Chapter 8.5
-
Update monitor : cf. Chapter 8.6
-
Shutting down : cf. Chapter 8.7
To note: The NIPE Monitoring can be used with a conscious patient as well as with an
unconscious one, whenever the clinician
needs it.
6
NIPE display
6.1
ECG capture
The lower sub-window displays the ECG acquired by the monitor (figure 8).This ECG is
filtered from all environmental artefacts.
Figure 8 - ECG
WARNING:
On the main screen,
always check the good quality of the ECG signal:
-
At the same time with the signal morphology and the scrolling in the ECG scope.
-
In the bottom right hand corner of the screen :
-
Good quality signal
-
Medium quality signal
-
Poor quality signal
Case n°1: No signal is displayed: