CUBEScan
TM
BioCon-900S Operator’s Manual
Page 54
TRANSMITTER
(W)
d = 1.2
√ 𝐏
d = 1.2
√ 𝐏
d = 2.3
√ 𝐏
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE
: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE
: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
9.2.5
Components conformance to standards
To maintain electromagnetic interference (EMI) within certified limits, the device must be used with the
cables, components specified or supplied by Mcube Technology. The use of cables or compoments other
than those specified or supplied may result in increased emissions or decreased immunity of the system.
EMC standards
Compoment
Max length
Power Cord
1.83 m(6ft)
Adapter
1 m (3.3ft)
9.3
EXCESSIVE TEMPERATURE TEST RESULT
Transducer type →
External use
Test to be applied ↓
Simulated use test
Temperature rise
The ambient temperature :
22.8
℃
The temperature rise to:
24.6
℃
Still air test
(no gel)
Temperature rise
The ambient temperature :
23.1
℃
The temperature rise to:
25.5
℃
9.4
SPECIFICATIONS OF COMPONENTS
9.4.1
WIRELESS CHARGING
Main Chipset
TI (BQ500210RGZ)
Coil type
1 Coil (A10)