CUBEScan
TM
BioCon-900S Operator’s Manual
Page 48
8.9
CALIBRATION
BioCon-900S must be calibrated every 12 months to ensure accurate results. Calibration ensures accurate
alignment of BioCon-
900S’s internal system. You can take scan, even if calibration is not performed by the
designated date. But result can be compromised. There are following methods for the calibration.
CAUTION
Calibrate BioCon-900S when the battery is charged over 60%.
8.9.1
PHANTOM CALIBRATION
Phantom Calibration
1)
Place the CubeScan phantom on a flat
surface and open the cover of the phantom.
2)
Pour about 5ml of water or more on the center
surface of the phantom
3)
Place the phantom cover on the top of the
CubeScan phantom. Check if the cover is in a
stable and flat position.
4)
Put the probe head into the probe holder firmly.
5)
Turn the probe on and Tap the Setup icon (
)
on the bottom right side of the Home screen.
6)
Tap the Next icon (
) on the Setup screen
and select Phantom Calibration icon (
)
7)
Tap the Scan icon (
) to start the calibration.
You can cease the phantom calibration by
tapping the Cease icon ( ). Tap the Exit icon
(
) to return to Setup screen, if you want.
8)
The calibration progress bar appears, when the
calibration is started. The progress bar indicates
the progress of calibration. It takes about
3minutes to complete the calibration. When the
calibration is completed, 100% sign is displayed
on the screen.
9)
Tap the Ok icon (
) and the latest calibration
date is updated. Tap the Exit icon (
) to return
to the Setup screen.