CUBEScan
TM
BioCon-900S Operator’s Manual
Page 49
8.9.2
CALKIT CALIBRATION
Calkit Calibration
1)
Open the cover of the calibration kit and pour the
saline solution.
Ideal calibration condition
You can pure water instead of the saline solution.
Ensure that water does not contain any air bubbles
before starting the calibration.
2)
There is a mark on the inside of the Calkit
showing the
Water Level
. The
Water Level
is
about 1cm below the top edge. Pour water to the
Water Level.
3)
Close the cover of the calibration Kit. Make sure
that the arrow of the Calkit cover is aligned with
the nipple of the Calkit body.
4)
Align the probe’s Scan Button
with the arrow
of the cover and put the probe head firmly into
the cover.
5)
Turn the probe on and tap the Setup icon (
)
on the botton right side of the Home screen.
6)
Tap the
Next
icon (
) on the Setup screen and
select the
CalKit Calibration
icon (
)
7)
Tap the Scan icon (
) to start the calibration.
You can cease the calkit calibration by tapping
the Cease icon ( ). Tap the Exit icon (
) to
return to the Setup screen, if you want.
8)
The calibration progress bar appears, when the
calibration is started. The progress bar indicates
the progress of calibration. It takes about
3minutes to complete the calibration. When the
calibration is completed, 100% sign wil be
displayed on the screen.
Tap the
Ok
icon (
) and the latest calibration
date is updated. Tap the
Exit
icon (
) to return
to the Setup screen.
8.9.3
CALIBRATION REMINDER
1)
Calibration Reminder Dialog box
Calibration Reminder Dialog box appears on the top screen under the conditions as follows:
-
After 1 year since the last calibration
Tap the Ok icon (
) to return to the Home
screen. The Calibration Reminder Dialog box disappears.
2)
Calibration Reminder icon
The
Calibration Reminder
icon (
) is displayed on the Home screen after the Calibration Reminder