![Mazet Sante Biostim 1.0 User Manual Download Page 7](http://html1.mh-extra.com/html/mazet-sante/biostim-1-0/biostim-1-0_user-manual_1752024007.webp)
Page 7 on 42
2022/09/27
MEG010XN102-A11 (ENG)
2.2
Technical specifications
2.2.1
General characteristics
•
Operating temperature: 15°C to 40°C.
•
Storage temperature: -20°C to 70°C.
•
Operating relative humidity: 30% to 75%.
•
Operating altitude: < 2000 metres
2.2.2
Technical characteristics of the device
•
Case dimensions:
33.7 x 28 x 6.7 cm
•
Case weight: 3
.1 Kg
•
Housing colour:
white
•
Power supply:
110-230VAC - 50-60Hz
•
Power consumption:
55VA
max
•
Fuses: 2x size 5x20mm -
T1.25AH-250V
•
Class I
electrical equipment
•
Medical
Class IIa
equipment.
•
Applied part
type BF
•
Liquid protection type
IPX0
.
•
Communication with the PC: Optically isolated
USB.
•
Power on indication by a green light on the front panel.
•
Possibility to stop the stimulation by an emergency stop bulb.
•
1 or 2 Electro channels. Each channel has the following features:
•
Current generator :
-Output currents of each generator adjustable from
0 to 100mA
(+/-10%).
-Under a load impedance of 1k
Ω
(or more), at max current, the voltage is
limited to
100V
-20%/+10% (peak value).
-Under a load impedance of less than 1k
Ω
, the voltage level is limited according to the
impedance (10volts for 100
Ω
, 50volts for 500
Ω
)
-If the impedance is too high (above 10 k
Ω
), the current can be cut off:
unstuck
electrode
function
The rectangular shaped signals are biphasic (symmetrical pulses with zero average), the
pulse width is adjustable from
50µs to 10ms
, the frequency is adjustable from
1Hz to
5kHz
.
The generators are electrically independent (no current flows between the 2 electrodes of
the 2 generators).
Indication of the activation status of the output by a yellow LED.
•
Biofeedback activity measurement: Full scale sensitivity: 2mV (peak-to-peak)
•
0, 1 or 2 channel Biofeedback pressure
sensitivity range:
400 mBar