1 Introduction
1-5
Matsusada Precision
AK series
・
If the Load frame is grounded
(
1-2kV output models
)
・
If the Load frame is grounded
(
over 3kV output models
)
● Make sure to connect the GROUND terminal correctly as shown in Fig. 1.2, Fig. 1.3, Fig. 1.4 and
Fig. 1.5.
Poor grounding is dangerous and may cause an electric shock or damage the power supply.
● When load short-circuit or discharge is forecasted, make the GROUND cable thicker and shorter.
Ground at one point
Fig. 1.4
Load
Load frame
Rack
– mount case
Wires should be :
・
more than 5SQ
・
less than about 2m
・
not bared
MUST
MUST
Black cable
(
M
6)
Output cable
Ground terminal
Ground at one point
Fig. 1.5
Load
Output cable
Ground terminal
Load frame
Rack
– mount case
MUST
MUST
Wires should be :
・
more than 5SQ
・
less than about 2m
・
not bared