IT6300B user manual
Track (sync output setting)
This function configures the instrument for tracking operations of CH1 and CH2. Possible
Press
button to confirm your
In tracking mode,once the parameters of any one channel are changed,other channels
will change proportionally.
For example, set voltage and current of CH1 and CH2 as follows, CH1: 4V, 1A
;
CH2: 8V,
2A. Press (Shift) +
display as follows:
(
Menu
)
into Menu, and press to select Comb, VFD will
CH1+CH2
Select “CH1+CH2” and press
to confirm.The VFD will display
“
Track Set
Success!” and escape this screen after 2S..
Front display as follows in condition of output off and meter state.
For example:In setting status, if voltage of CH1 is set as 2V, voltage of CH2 will
automatically change to 4V
proportionally.
Notice:Tacking function is disabled to the channel with 0V or 0A setting.In the
former example,if CH2 setting is 0V or 0A,then when CH1 voltage is adjusted to
2V,CH2 will remain unchanged.
Parameters in Serial,Parallel or Tracking mode
Maximum voltage
Operate “CH1+CH2” in series,the max voltage is 62V the sum of the max voltage
of CH1 and CH2.
29
USER MANUAL
Button to confirm your set, and press ESC to quit the operation.