20
Navigate further into the program options by using the ⁄ down
arrow key.
LOADED PROGRAMS
One of the available programs is always loaded into the
memory for use. How do you verify which program is loaded
for operation of the Composite Sampler? To verify which
program is loaded for operation of the unit, access the menu
by pressing the MENU key and selecting VIEW/EDIT
PROGRAMS, then press the ENTER key.
The display will change to show 2 program options. One of
these options will show a darkened or filled in box next to the
program name. This is the program which is loaded for operation.
See the example below.
In this case PROGRAM #3 is loaded for operation as shown
in the above example.
CONTINUOUS PUMPING
This program is for fluid transfer in the field. The pump will
start when the START/STOP button is pressed or the REMOTE
CONTACT CLOSURE is closed. The pump will stop when
the same button is pressed again, the REMOTE CONTACT
CLOSURE opens or the fluid level sensor in the sample
reservoir causes the system to stop because the reservoir
is full. This program cannot be modified in any manner.
This program selection is not visible when you first access
the program menu. Select VIEW/EDIT PROGRAMS and press
ENTER. The display will change to read as follows.
The arrow at the far left of the display on line 1 shows a
bi-directional arrow, indicating there is an available option
above PROGRAM #1.
Use the up arrow key to navigate upward in the menu. The
display will change to read:
You may now select CONTINUOUS PUMPING by pressing
the ENTER key.
c
PROGRAM #2
c
PROGRAM #3
c
PROGRAM #3
c
PROGRAM #4
c
PROGRAM #4
c
PROGRAM #5
g
PROGRAM #3
c
PROGRAM #4
g
PROGRAM #1
c
PROGRAM #2
g
CONTINUOUS PUMPING
c
PROGRAM #1