HYS006602
07
06
Using this appliance in an environment
with a strong radiated radio-frequency
electromagnetic field (approximately
3V/m) may influence its measuring
accuracy. The measuring result can be
strongly deviating from the actual value.
5. DC Current Measurement
5.1
5.2
5.3
Set the AC/DC current transducer function
switch at desired A position.
Set a DMM at DCV function and a
relevant range.
Connect the output plug of AC/DC current
transducer with a DMM and combined
firmly.
5.5
5.6
Adjusting the DCA ZERO knob until the
DMM display show “0”, when does not
measurement before.
Press the trigger to open transformer jaws
and clamp one conductor only, making sure
that the jaws are firmly closed around the
conductor.
Read the reading on the LCD display of
DMM along with the polarity of the red
lead connection.
5.4
6. AC Current Measurement
6.1
6.2
6.3
Set the AC/DC current transducer function
switch at desired A~ position.
Set a DMM at DCV function and a relevant
range.
Connect the output plug of AC/DC current
transducer with a DMM and combined
firmly.
Press the trigger to open transformer jaws
and clamp one conductor only, making
sure that the jaws are firmly closed around
the conductor.
6.4
6.5
Read the reading from the digital display
of the DMM.
7. Maintenance
To keep the instrument clean, wipe the case with a damp
cloth and detergent, do not use abrasives or solvents.
Any adjustment, maintenance and repair shall be
conducted by service personnel.
Accessories
•
Operator's instruction manual
•
Gift box
•
9 volt battery. NEDA 1604 6F22 006P type
05
04
3. Specifications
Accuracy is specified for a period of one year after
calibration and at 18°C to 28°C(64F to 82F) with relative
humidity to 80.
Accuracy specifications are given as:
±of reading digits
This instrument must not be used on
uninsulated conductors at a voltage
greater than 250V ac rms. or 250V dc
3.1 General
DC CURRENT RANGE: 1A dc to 1000A dc.
AC CURRENT RANGE: 1A ac to 1000A ac. RMS
OUTPUT VOLTAGE: 1mV dc. per 1 A dc or 1A ac.
WORKING VOLTAGE:
1000V CAT.II Per IEC 1010 - 1
MAXIMUM ALTITUDE: 2000m
OPERATING TEMPERATURE:
0°C to 40°<C 80 relative humidity
STORAGE TEMPERATURE:-10°C to 50°C
TEMPERATURE COEFFICIENT:
0.2×(Spec Acc'y) /°C <18°C or>28°C
MAXIMUM JAW OPENING: 55mm
SIZE: 256×104×47
WEIGHT: Approx. 520g
3.2 Electrical specification dc current
Range
Measure
Output
Accuracy
1000mV
1000A
±
±
3.0% 5
A
Overload Protection:
1200A for 60 seconds maximum.
DC Current
AC Current
Range
Measure
Output
Accuracy
1000mV
1000A
±
±
3.0% 5
A
Overload Protection:
1200A for 60 seconds maximum.
Frequency range: 50Hz to 60Hz.
4. Precaution and
Preparations for Measurement
4.1
4.2
4.3
Dot not applies the voltage to the output
plugs of AC/DC current transducer.
Do not use or store this instrument in a high
temperature or high humidity environment
and do not store the unit indirect sunlight.
Do not measure current before the AC/DC
current transducer is not combined with
DMM.
AC / DC CURRENT TRANSDUCER
AC / DC CURRENT TRANSDUCER
AC / DC CURRENT TRANSDUCER
AC / DC CURRENT TRANSDUCER