Using this appliance in an environment 
with a strong radiated radio-frequency 
electromagnetic field (approximately 
3V/m) may influence its measuring
accuracy. The measuring result can be
strongly deviating from the actual value.

5. DC Current Measurement

5.1

5.2

5.3

          Set the AC/DC current transducer function
          switch at desired A   position.
          Set a DMM at DCV function and a
          relevant range.

          Connect the output plug of AC/DC current
          transducer with a DMM and combined 
          firmly.

5.5

5.6

          Adjusting the DCA ZERO knob until the 
          DMM display show “0”, when does not 
          measurement before.

          Press the trigger to open transformer jaws 
          and clamp one conductor only, making sure
          that the jaws are firmly closed around the
          conductor.

          Read the reading on the LCD display of
          DMM along with the polarity of the red
          lead connection.

5.4

6. AC Current Measurement

6.1

6.2

6.3

          Set the AC/DC current transducer function 
          switch at desired A~ position.

          Set a DMM at DCV function and a relevant
          range.

          Connect the output plug of AC/DC current
          transducer with a DMM and combined 
          firmly.

          Press the trigger to open transformer jaws
          and clamp one conductor only, making 
          sure that the jaws are firmly closed around
          the conductor. 

6.4

6.5

         Read the reading from the digital display 

     

of the DMM.

7. Maintenance

To keep the instrument clean, wipe the case with a damp
cloth and detergent, do not use abrasives or solvents.
Any adjustment, maintenance and repair shall be
conducted by service personnel.

Accessories

   

 Operator's instruction manual

  

 Gift box

  

 9 volt battery. NEDA 1604 6F22 006P type

05

04

3. Specifications

Accuracy is specified for a period of one year after 
calibration and at 18°C to 28°C(64F to 82F) with relative
humidity to 80.
Accuracy specifications are given as:
±of reading  digits

This instrument must not be used on 
uninsulated conductors at a voltage
greater than 250V ac rms. or 250V dc

3.1 General

DC CURRENT RANGE: 1A dc to 1000A dc.
AC CURRENT RANGE: 1A ac to 1000A ac. RMS
OUTPUT VOLTAGE: 1mV dc. per 1 A dc or 1A ac.
WORKING VOLTAGE:
1000V CAT.II Per IEC 1010 - 1
MAXIMUM ALTITUDE: 2000m
OPERATING TEMPERATURE:
0°C to 40°<C 80  relative humidity
STORAGE TEMPERATURE:-10°C to 50°C
TEMPERATURE COEFFICIENT:
0.2×(Spec Acc'y) /°C <18°C or>28°C
MAXIMUM JAW OPENING: 55mm
SIZE:  256×104×47
WEIGHT: Approx. 520g 

3.2 Electrical specification dc current

Range

Measure

Output

Accuracy

1000mV

1000A

±

±

3.0% 5

A

Overload Protection: 
1200A for 60 seconds maximum.

DC Current

AC Current

Range

Measure

Output

Accuracy

1000mV

1000A

±

±

3.0% 5

A

Overload Protection:
1200A for 60 seconds maximum.
Frequency range: 50Hz to 60Hz.

4. Precaution and 
    Preparations for Measurement

4.1

4.2

4.3

          Dot not applies the voltage to the output 
          plugs of AC/DC current transducer.

          Do not use or store this instrument in a high
          temperature or high humidity environment 
          and do not store the unit indirect sunlight.

          Do not measure current before the AC/DC
          current transducer is not combined with
          DMM.