background image

1.2 Line regulation :     CV

0.02%+3mV                        CC

0.5%+3mA  

1.3 Load regulation :    CV

0.02%+10mV(I

6A)          CC

0.5%+10mA(I

6A) 

                                              CV

0.1%+5mV(I>6A)                CC

0.5%+20mA(I>6A) 

1.4 Protection : constant current and short-circuit protection 

1.5 Voltage indication accuracy :    LED

±

1%+2digits, analogue display 2.5% 

1.6 Current indication accuracy :    LED

±

2%+2digits, analogue display 2.5% 

1.7 Environment : 0 ~+40 

0

C      relative humidity : <90% 

2 OPERATION   

2.1 Front panel controls. 

(1) Current indication. 

(2) Voltage. Indication.   

(3) Fine adjustment of current 

(4) Coarse adjustment of current. 

(5) Fine adjustment of output voltage. 

(6) Coarse adjustment of output voltage. 

(7) Power ON/OFF switch 

(8) Negative output terminal.   

(9) Ground connection.   

                                                                                                    -

(10) Positive output terminal 

    - 

2

Summary of Contents for HY3010E

Page 1: ...USER S MANUAL DC REGULATED POWER SUPPLY ...

Page 2: ...050E HY5005E HY5010E HY5020E HY10010E Regulated Output Voltage 0 30V 0 30V 0 30V 0 30V 0 50V 0 50V 0 50V 0 100V Regulated Output Current 0 10A 0 20A 0 30A 0 50A 0 5A 0 10A 0 20A 0 10A Display Double LED Double LED Double LED Double LED Double LED Double LED Double LED Double LED Ripple and Noises 0 5mVrms 1 0mVrms 3 0mVrms 3 0mVrms 0 5mVrms 1 0mVrms 3 0mVrms 10mVrms 1 TECHNICAL PARAMETER 1 1 Input...

Page 3: ...logue display 2 5 1 6 Current indication accuracy LED 2 2digits analogue display 2 5 1 7 Environment 0 40 0 C relative humidity 90 2 OPERATION 2 1 Front panel controls 1 Current indication 2 Voltage Indication 3 Fine adjustment of current 4 Coarse adjustment of current 5 Fine adjustment of output voltage 6 Coarse adjustment of output voltage 7 Power ON OFF switch 8 Negative output terminal 9 Groun...

Page 4: ... 7 and connect the load to the output terminals 8 10 Adjust controls 3 4 to set the desired output current 2 2 3 For restricted current protection mode switch on the power ON OFF switch 7 adjust controls 3 4 anti clockwise to the minimum position adjust controls 5 6 clockwise to set the desired output voltage level and then connect the load to the output terminals 8 10 Adjust controls 3 4 clockwis...

Page 5: ...rvicing and servicing should be referred to a qualified person 3 3 The unit should be stored in a dry and well ventilated place and the power cord removed if storing for long periods 4 ACCESSORIES 4 1 Power cord one piece 4 2 Instruction manual one piece ...