SETUP

Setup is used to view or change the instrument parameters.

To enter setup press SHIFT+SETUP when the meter is in measure-
ment mode.

"Set" is displayed on the upper LCD. The lower LCD displays the blinking code of the
current setup item.

Select the desired setup item using the 

 or 

 key.

Press SHIFT+CFM to confirm.

Note:

Note:

Note:

Note:

Note: If SHIFT+SETUP are pressed before item confirmation, the meter will escape and

return to measurement mode.

Once the desired setup item has been selected, its current value blinks (if it is a
changeable parameter).

Summary of Contents for MI 306

Page 1: ...C TDS Meter EC TDS Meter EC TDS Meter EC TDS Meter EC TDS Meter Mi 306 A A A A Automatic L utomatic L utomatic L utomatic L utomatic Logging EC TDS Meter ogging EC TDS Meter ogging EC TDS Meter ogging...

Page 2: ...LOGGING FUNCTION 14 DATA TRANSFER TO PC 17 BATTERY REPLACEMENT 18 CONDUCTIVITY VS TEMPERATURE CHART 18 ACCESSORIES 19 DISPLAY DISPLAY DISPLAY DISPLAY DISPLAY A PRIMARY DISPLAY B AUTO ENDPOINT MODE IND...

Page 3: ...or with SHIFT to enter setup mode 5 ATC TC key to select temperature compensation mode or with SHIFT to view the temperature coefficient value 6 MEM MR key to store or with SHIFT recall measurements 7...

Page 4: ...ly ATC or manually MTC compensated for temperature The temperature coefficient value is user selectable It is possible to disable the temperature compensation and measure the actual conductivity NOTC...

Page 5: ...0 1 S cm from 30 0 to 299 9 S cm 1 S cm from 300 to 2999 S cm 0 01 mS cm from 3 00 to 29 99 mS cm 0 1 mS cm over 30 0 mS cm TDS 0 01 mg L from 0 00 to 14 99 mg L 0 1 mg L from 15 0 to 149 9 mg L 1 mg...

Page 6: ...aCl Calibration 1 point with HI 7037 buffer optional Temperature 2 points at 0 and 50 C Calibration plus 1 C adjustment Temperature Automatic or Manual from 0 to 60 C Compensation can be disabled to m...

Page 7: ...upper LCD The lower LCD displays the blinking code of the current setup item Select the desired setup item using the or key Press SHIFT CFM to confirm Note Note Note Note Note If SHIFT SETUP are press...

Page 8: ...to 2098 1999 dAT Date dd mm 01 01 hou Hour hh mm 00 00 uEr Firmware release Chr Battery level test To change the value use the or key Press SHIFT CFM to confirm Note Note Note Note Note Press SHIFT S...

Page 9: ...sired range EC TDS NaCl is selected on the LCD Allow for the reading to stabilize The upper LCD displays the measure in the selected range while the temperature is displayed on the lower LCD Notes Not...

Page 10: ...setup mode and selecting the tdS item see SETUP section Always set the reference temperature to 25 C when measuring TDS 7 When the use of an alternate function MR SETUP CFM FIXED TC and HOLD is reques...

Page 11: ...the range options TEMPERATURE COMPENSATION Three options of compensating temperature are available Automatic Atc The probe has a built in temperature sensor the value of the tempera ture is used to a...

Page 12: ...S cm 12 88 mS cm 80 0 mS cm 111 8 mS cm To enter EC calibration select the EC range and press the CAL key Note TDS reading is automatically derived from the EC reading and no specific calibration for...

Page 13: ...the MA 9066 MA 9066 MA 9066 MA 9066 MA 9066 calibration solution sea water solution as a 100 NaCl standard solution To enter NaCl calibration select the NaCl from RANGE key and press CAL Rinse the pro...

Page 14: ...and the MEM key is pressed the sample will not be stored and the LCD will display FULL In this case it is necessary to delete some data from memory to proceed To view logged data To view logged data...

Page 15: ...o skip to another sample using the up and down arrow keys For example if the reading of a sample is displayed pressing the up arrow key will cause the meter to display the reading of the next sample A...

Page 16: ...Press SHIFT HOLD to escape without data deletion When viewing through the logged data the NULL message will be displayed when selecting a deleted sample To delete all data in memory To delete all dat...

Page 17: ...cable start the terminal program and set the communication options as follows 8 N 1 no flow control To send a command to the meter the scheme is command CR The computer sends the command expressed as...

Page 18: ...es with increasing temperature It is affected by the type and number of ions in the solution and by the viscosity of the solution itself Both parameters are temperature dependent The depen dency of co...

Page 19: ...t defects in materials and manufacturing for a period of 3 years from the date of purchase Electrodes and Probes are warranted for 6 months If during this period the repair or replacement of parts is...

Page 20: ...ls kik t sor 11 6726 Szeged Hungary Tel 36 62 428 050 Fax 36 62 428 051 e mail sales milwaukeeinst com Milwaukee Instruments Inc 2950 Business Park Drive Rocky Mount NC 27804 USA Tel 1 252 443 3630 Fa...

Reviews: