cooling slots or cover joints, it is advisable to wrap
metal foil tightly around the appliance and connect
the test probe to the metal foil then perform Class II
insulation tests.
Testing in Manual Mode
u
Press SHIFT and INS II (CLASS II) to select the Class
II insulation test and display the diagram screen.
u
Press OK for the HPAT600 to display the test screen
and perform the Class II insulation measurement.
If ‘FAIL’ is displayed after the Class II insulation
measurement, refer to 3.12. (Earth Continuity/Insulation
Failure)
Testing in Auto Mode
u
Press SHIFT and INS II (CLASS II) to select the
Class II insulation test and display the test screen.
u
The HPAT600 will perform the Class II insulation
measurement.
If ‘FAIL’ is displayed after the Class II insulation
measurement, refer to 3.12. (Earth Continuity/Insulation
Failure)
Class II Test
Connections
30