Class ITest
Connection
Testing in Manual Mode
u
Press
CLASS I
to select the Class I appliance test and
display the diagram screen.
u
Press
OK
for the HPAT600 to display the VISUAL
CHECK screen.
u
If any visual or fuse checks fail, select
Fail
and press
OK
to abort the test and return to the select test
screen.
u
If all visual and fuse checks pass, select
Pass all
and
press
OK
for the HPAT600 to display the test screen
and perform the fuse check.
If
‘FAIL’
is displayed after the fuse check, refer to 3.11.
(Fuse Check/Simulated Load Failure)
u
Press
OK
for the HPAT600 to perform the Class I
earth continuity measurement.
u
If
‘FAIL’
is displayed after the Class I earth continuity
measurement, refer to 3.12. (Earth Continuity/
Insulation Failure)
u
Press
OK
for the HPAT600 to perform the Class I
insulation measurement.
u
If
‘FAIL’
is displayed after the Class I insulation
measurement, refer to 3.12. (Earth Continuity/
Insulation Failure)
21