background image

OPERATOR’S INSTRUCTIONS 

Key description 

 

Tare

   

 

 

- Set the display to zero by storing the current weight in 

the tare memory 

  

 

On 

Off

 

 

 

- Turns the scale on or off, 

Units 

 

 

 

- Select kilograms, pounds or ounces 

 

   

Net / Gross

 

 

- Select either Gross weight or Net weight when   

      

 

 

 

 

tare 

weight 

has 

been 

entered 

 

   

Count

  

 

 

- Count how many times you weigh 

o weigh on the scale turn the scale on using the 

On/Off

 key.    After the display is at zero place an item 

to be weighed on the platform.    The display will show the weight.    The unit will be shown in 
kilograms unless the user has selected either pounds or ounces previously. 

If an empty container is used the weight can be ignored by zeroing the display.    To zero the display 
presses the 

TARE

 key.    Then when material is placed in the container only the weight of the material 

will be shown. 
Switch between showing the Net Weight (Material Only) and the Gross Weight (container plus material) 
by pressing the 

Gross/Net

 key. 

 

D I S P L A Y   S Y M B O L S  

Symbol Description 

Î{Í

  

scale in auto zero range    (indication = accurate zero) 

{

 

result is stable   

kg , lb or oz 

scale in weighing mode 

Lo-bAt 

battery is weak, after 5 minutes the scale will turn off   

 

Show the Net Weight is being display (NW) or the Gross Weight is 

 displayed (GW) 

pcs 

You can select the sample size from 10, 25, 50 or 100 pieces 

CALIBRATION 

 

Occasionally the scale should be verified that it is weighing correctly by measuring a known mass of 
approximately the scale capacity.      Tare the scale, place the mass on the center of the platform and 
note the reading.    Calibrate the scale if necessary.   
Before beginning make sure you have the mass specified and you know the mass is accurate.   

PROCEDURE 

Turn the scale off. 
Press and hold the 

On/Off

 key then press the 

Units

 key 3 times. Release the 

ON/OFF

 key. The 

display will show a value from the internal counters.    The typical value is 2000 to 16000. 
Press the 

Tare

 key.    Display will show zero. 

Place the calibration weight (below) on the scale, centers on the platform. 

 

  MPS  30   

20kg mass 

 

  MPS  60   

40kg mass 

MPS 150   

100kg mass 

Press the 

Tare

 key again.    The display will show “ CAL” then returns to weighing showing the value of 

the weight used.    Remove the weight.    Verify the scale is calibrated correctly. 

 
 

Summary of Contents for Boxer MPS 30

Page 1: ... a regulated 6 to 15VDC 100 mA or using batteries inside of scale s head 6 x AA Size Remove the cover from the battery compartment under the stainless steel pan by lifting the tab Install six AA size batteries as shown It is suggested Alkaline batteries be used for best life SPECIFICATIONS Model MPS 30 MPS 60 MPS 150 Maximum Capacity 30 kg 60 kg 150 kg Readability 0 01 kg 0 02 kg 0 05 kg Resolutio...

Page 2: ...zero range indication accurate zero result is stable kg lb or oz scale in weighing mode Lo bAt battery is weak after 5 minutes the scale will turn off Show the Net Weight is being display NW or the Gross Weight is displayed GW pcs You can select the sample size from 10 25 50 or 100 pieces CALIBRATION Occasionally the scale should be verified that it is weighing correctly by measuring a known mass ...

Page 3: ...nit Press the Unit key to cycle through the options Unit A Off CAP dIuS End SEt Service Only At each option press the Net Gross key to enter the functions and then press the Unit key to see the options pressing the Net Gross key again to return to the functions menu Each weighing unit can be enabled or disabled The Automatic Turn off time can be set or disabled off The other functions do not apply...

Page 4: ...tandby alternately Minimum sample weight is 20g Manufacturer s Declaration of Conformity This product has been manufactured in accordance with the harmonised European standards following the provisions of the below stated directives Electro Magnetic Compatibility Directive 89 336 EEC Low Voltage Directive 73 23 EEC MARSDEN Equipment Co Ltd 4 ...

Reviews: