Marquette Hellige GmbH
MicroSmart V 2.xx
Page 58
Servicing Instructions 227 470 35 B - 97.12
Electrical Diagram for Patient Leakage Current Test
For protection of the test person the following values of resistor R may be used:
Typ BF
22 kOhm (120 to 130 V)
47 kOhm (220 to 240 V)
Typ CF
100 kOhm (220 to 240 V)
8.3 Maintenance, cleaning, disinfection
MicroSmart maintenance, cleaning, disinfection is performed in accordance with the MicroSmart
or MircoSmart MC, User’s Manual, Chapter 12, "Cleaning, disinfection and maintenance", as
applicable.
Applications requiring extensive recordings may result in deposits on the thermal array printhead
which normally do not, however, have any adverse effect on the printing quality. This can be
removed with a soft, fluff-free cloth soaked in an alcohol-based cleaning agent (e.g., surgical
spirit).