Servicing Instructions 227 470 35 B - 97.12
4.8 Additional functions for final test
A number of additional functions can be selected in the test mode by striking certain key
combinations. These functions are primarily intended for final testing. This is why they are not
explicitly available as selectable items within the test mode. Instead, they can be reached via
certain key combinations. These are:
-
setting time and date
-
reading in the serial number
Setting time and date
This mode can be reached from the selection level for appliance tests with the key combination
"Man + Config".
Use the "-->" key to select the setting for the date, and the key "<--" to select the setting for the
time. The setting steps which follow are designed for an automated setting. Instead of showing
the current data, the unit will display the setting returned to "zero".
Reading in the serial number for determining the option key
(MicroSmart MC only)
Necessary equipment for reading in the serial number:
Hardware
-
PC or Laptop
-
Interface Cable PC - Irda Adapter (Partnumber 223 426 01)
-
Irda Adapter (Partnumber 930 117 72)
Software
-
use Terminal-Program (Windows 3.11) or
Hyperterminal-Program (Windows 95and Windows NT)
Transfer parameters: 9,600 baud rate, 1 stop bit, no parity, 8 data bits
MicroSmart MC
The serial number is read in via the Irda interface. Starting from the selection level for the
appliance test, the read-in function for the serial number is reached via the key combination
"HR^ + setup". The unit immediately moves into the receiving mode and reads in the incoming
data. Only numerical characters (maximum 9) will be accepted. The transferred serial number
appears on the display. If no serial number is transferred, an ERROR-message (Receive error)
appears on the display.
If no serial number is entered, the unit will show an appropriate 5 second visual and acoustic
signal each time the unit is switched on.
PC or Laptop
Connect the Interface Cable to a free COM port. Connect the Irda Adapter to the Cable. Start the