3.

 

Features 

The Tester is a microcomputer controlled, High Voltage Insulation Resistance 
Tester with 4-range for measuring insulation resistance and AC/DC Voltage, 
Resistance/ Continuity test. 
 

 

Designed to following safety standards: 

IEC 61010-1 (CAT IV 600V Pollution degree 2) 
IEC 61010-031 (Requirements for hand-held probes) 

 

Insulation test range: 0.1MὨ to 6kὨ 

 

With auto-discharge function & voltage output warning function. 

 
When insulation resistance like a capacitive load is measured, electric charges 
stored in capacitive circuits are automatically discharge after measuring. 
Discharge can be checked with live voltage graph. 

 

Backlight levels selector function to facilitate working at dimly illuminated 

location or at nighttime work. 

 

Live circuit warning symbols plus audible warning. 

 

Live circuit detection prevents insulation test if voltage >30V is detected for 

added user protection 

 

With auto-power off function and battery check. 

 

With Timer measurement function 

Automatically performs a measurement at the set time. 

 

With polarization index measurement (PI) 

The polarization index can be measured by the automatic measurement function 
of the ratio of resistance in 1 min. and 10 min. two point time and automatic 
(Pass/Fail). 

 

With Dielectric absorption ratio measurement (DAR) 

The dielectric absorption ratio measured by the automatic measurement function 
of the ratio of resistance in 30sec. and 1min. two point time and automatic 
(Pass/Fail). 

 

Auto ranging with large 6000 counts LCD display with bargraph. 

 

MAX/MIN, PEAK, Relative value & Data hold functions for DC/AC voltage 

measurements. 

 

Power source 1.5V x 8 (LR14/R14) Batteries. 

 

Summary of Contents for MIR-505

Page 1: ...HIGH VOLTAGE INSULATION TESTER Product Code MIR 505 USER MANUAL WWW MARMONIX CO HIGH VOLTAGE INSULATION TESTER...

Page 2: ...rement Continuity Check D Measuring Insulation Resistance 9 Auto discharge function 10 Principal of Insulation Resistance Measurement A Continuous Insulation Resistance Measurement B Timer Insulation...

Page 3: ...ory II Equipment of Overvoltage Category II is energy consuming equipment to be supplied from the fixed installation Note Examples include household office and laboratory appliances Overvoltage Catego...

Page 4: ...Resistance tester with ant parts or cover removed When carrying out insulation measurement do not come in contact with the circuit test Warning Do not use the Insulation resistance tester if it is da...

Page 5: ...When the instrument will not be in use for a long period place it in storage after removing the batteries Symbols Risk of electric shock Equipment protected by double or reinforced insulation DC measu...

Page 6: ...or at nighttime work Live circuit warning symbols plus audible warning Live circuit detection prevents insulation test if voltage 30V is detected for added user protection With auto power off functio...

Page 7: ...load 1 to 1 2mA at 1M load 1 to 1 2mA at 2 5M load 1 to 1 2mA at 5M load Recommended use of Power Adaptor Short Circuit Current Approx 1mA Accuracy 2 5 reading 15digits at 0 005 to 600 0M 3 reading 15...

Page 8: ...ts 601 to 6 00k 0 001k 1 5 reading 15digits Continuity Buzzer 50 Short Circuit Current 200mA Open Circuit Test Voltage 4 5V DC AC Voltage Measurement Range Resolution Accuracy DC Voltage 0 5 to 600 0V...

Page 9: ...egory II Pollution Degree 2 Altitude up to 2000 meters Indoor Use only Power source 8 x 1 5V R14 batteries Operating conditions 0 to 40 C 32 to 104 F 80 RH Storage conditions 10 to 60 C 14 to 140 F 80...

Page 10: ...ce test emergency stop button 6 LCD display 7 DC 13 5V Power adapter Input Terminal 8 EARTH High resistance measurement input 9 V input terminal GUARD Grounding protection input terminal 10 COM High v...

Page 11: ...display reading Measurement Inputs Terminal Description Insulation resistance measurement LINE 500V 1000V 2500V 5000V range Insulation resistance measurement high voltage line output LINE COM two plug...

Page 12: ...utomatically at any set time Time down Set timer time range down This is a function to conduct a test automatically at any set time PI Polarization index measurement press 2sec Clear timer function DA...

Page 13: ...o not touch it A Measuring Voltage DANGER To avoid electric shock please do not test if voltage to ground AC DC is over 600V even if the line voltage is below 600V Under voltage measurement please not...

Page 14: ...Low Resistance Measurement continuity Check Warning Do not run this test unless ACV DCV 0 Do not use this mode to check diodes 1 Set the Rotary switch to the Position 2 Connect the red test lead to t...

Page 15: ...ESS TO TEST button is pressed To check the insulation of electric equipment s or electric circuits measure their insulation resistance with the instrument Be sure to check the voltage which can be app...

Page 16: ...ouch the circuit under test immediately after testing Capacitance stored in the circuit may cause electric shock Leave test leads connected to the circuit and never touch the circuit until the dischar...

Page 17: ...in to 1 min 3 Pressing and hold turning the PRESS TO SET button clockwise to perform a timer measurement of insulation resistance Then the button is locked and timer measurement can be performed 4 Aft...

Page 18: ...seconds NOTES The following for reference only According to the standard IEEE 43 2000 with respect to the motor and related equipment for trouble shooting it will be for the insulation resistance tes...

Page 19: ...o repair or replace parts 12 DAR Standard DAR value 1 6 or more 1 25 1 6 1 25 or less Tested material status Excellent PASS Fail 13 Use of Guard Terminal When measuring a cable leakage current flowing...

Page 20: ...y the present reading but will continue to update and store the max and min readings 5 To exit MAX MIN PEAK mode press and hold MAX MIN for 2 seconds or presses PEAK MAX MIN Button Only for AC DC volt...

Page 21: ...on Press the key to turn ON OFF and select backlight levels The backlight will automatically turn off in 0 seconds 15 Auto Power Off To reserve battery power the tester enters into the Sleep Mode and...

Page 22: ...ejoin the case bottom and battery compartment and reinstall the screw 17 The Use of Power Adaptor Make sure the meter is power off and insert the power adaptor to the input terminal It is highly recom...

Page 23: ...Service Periodically wipe the case with a damp cloth and mild detergent Do not use abrasive or solvents To clean the terminals with cotton bar with detergent as dirt or moisture in the terminals can...