a.
Locate the screw in the slot.
b.
Push the assembly gently back in the horizontal plane guided by the screw, taking care to align the
trap and stainless steel connector
MISALIGNMENT CAN CAUSE THE END OF THE QUARTZ TRAP TO SNAP
c.
Apply gentle steady pressure to push the trap into the sealing O-ring located inside the steel trap
connector. If the pneumatics assembly does slide easily onto the cold trap remove trap pneumatics
and check seating of the o-ring.
d.
Secure the Trap Pneumatics Assembly in place, taking care to hold the assembly in place until the
screw has been fully tightened.