The analog and the digital filters should be disabled or set to their 
minimum acceptable values for highest peak capture speed.

 

 

EXTERNAL TRIGGER MODE  

 

This mode of operation is useful for measuring electrical contact activation force 
as well as synchronization of multiple instruments for a "snap-shot" view of the 
applied force or torque. When this mode is enabled through the configuration 
menu (see Section 3), the MODE key will sequence through an additional state 
which is indicated  by the flashing "C" or "T" indicator or "CW" or "CCW" 

indicator, depending on the type of 
sensor used.

 

 
When in this mode, the instrument 
stops updating the display when the 
trigger signal is applied. It is possible 
to capture the reading with a normally 
open contact (high to low transition of 
the trigger signal) or a normally closed 
contact (low to high transition). 
 
The display will show the captured 

reading until ZERO or CLEAR is pressed if the "edge" mode is set. The "level" 
mode provides for the display to hold the reading only until the trigger signal 
returns to its original state.Please refer to the diagram for connection details 
and to Section 3 for configuration information. 

 

 
6 SET 

POINTS 

 
This feature is useful for tolerance checking (GO/NO GO) or alarm indication in 
process control applications. Two limits, high and low, are specified and stored 
in the non-volatile memory of the instrument and all readings are compared to 
these limits. The results of the comparisons are indicated through the three 
open-collector outputs provided on the 9-pin connector, thus providing "under", 
"in range" and "over" signaling. These outputs can be connected to indicators, 
buzzers or relays as required.  After the Set Point mode is enabled through the 
configuration menu (see Section 3), pressing the MODE key will sequence 

through an additional step indicated by 
"SP" on the display. To enter or change 
the values of the set points press 
ENTER. The high set point is displayed. 
Use the 

and 

 

keys to increase or 

decrease the value and use MODE for 
changing between tension and 
compression or CW and CCW. When the 
desired value is displayed, press ENTER 

USER - SUPPLIED

SWITCH UNDER TEST

47K

+5V

GND

TRIG

3

5

GAUGE

USER - SUPPLIED

40 ma MAX

AS ABOVE

SAME

2N3904

GAUGE

9

8

5

GND

4

OUT

_

+

24V MAX

LOAD

Summary of Contents for BGI

Page 1: ...DIGITAL FORCE TORQUE INDICATOR Model BGI User s Guide...

Page 2: ...ed For additional information or answers to your application questions contact us and our technical support and engineering teams will be eager to help you Thank you again for your purchase and happy...

Page 3: ...in parentheses For a detailed description of the secondary functions see Section 3 POWER ENTER Turns power on and off UNITS Selects units of measurement ZERO Zeroes any tare value up to the full capac...

Page 4: ...has several features with programmable options allowing many user specified choices In order to access the configuration menu perform the following 1 Turn off the indicator 2 Press and hold MODE 3 Tu...

Page 5: ...Set point outputs enabled bcd d Mitutoyo BCD output disabled bcd E Mitutoyo BCD output enabled nPOL Mitutoyo readings without polarity absolute value POL Mitutoyo readings with polarity positive for...

Page 6: ...gs A E Average mode enabled A d Average mode disabled dEL Initial delay prompt At Average time prompt trF Trigger force torque value prompt CAL Calibration See Section 10 4 FILTERS For maximum flexibi...

Page 7: ...trigger signal returns to its original state Please refer to the diagram for connection details and to Section 3 for configuration information 6 SET POINTS This feature is useful for tolerance checkin...

Page 8: ...The process of averaging begins as soon as the programmed trigger reading is reached and is indicated by a flashing A The conclusion of the test is indicated by an alternating A and the calculated val...

Page 9: ...e PKCW Peak Clockwise mode PKCCW Peak Counter Clockwise mode CLR Clear peaks start a new average or external trigger test Z Zero display and perform the CLR function Request the displayed reading C Re...

Page 10: ...ngs and features using the same abbreviations as the commands to set them Any detected errors are reported back by means of the following error codes 10 Illegal command 11 Not applicable e g SPHn comm...

Page 11: ...transmitted if DATA is pressed during the average computation while in the Average mode or during the input scanning in the External trigger mode Mitutoyo BCD This output is useful for connection to...

Page 12: ...eration In some cases the display will show nnnn or uuuu to indicate excessive or insufficient calibration weight This can be caused by incorrect weights tare weight of over 10 of the full capacity or...

Page 13: ...cts in workmanship and materials Mark 10 Corporation will at its option repair or replace or refund the purchased price of goods found to be defective This remedy shall be the buyer s sole and exclusi...

Page 14: ...standard line of products we can provide modifications and custom designs for OEM applications Our engineering team is eager to satisfy any special requirements Please contact us for further informati...

Reviews: