Water Monitoring System
Ver.5.0 – Revision March 2022
Page
45
of
110
2.
Take the syringe and attach the needle. Break the neck of the ampoule – fill the syringe
with the entire content.
3.
Slowly inject the content into the PAH calibration chamber, taking care that no air is
mixed into the validation liquid. Look into the chamber to check and remove air
bubbles.
4.
Place the cleaned PAH sensor in the calibration chamber in an upright position. Wait for
the display of PAH value to approach the value of the used ampoule. If the displayed
value is within the tolerance in Table 6-2, the validation is successful, and a calibration
is not required. If the validation is not successful, refer to Figure 6-2.
5.
Empty out the validation liquid from the calibration chamber and wipe it dry with a new
cleaning pad. Clean the PAH sensor.
6.4
Calibration of the PAH Module
ATTENTION
Do not leave the sensor in the PAH calibration liquid more than 15 seconds since UV light
degrades the PAH calibration liquid.
Note
G6110:
Normally all local calibrations are done using the PAH calibration chamber
with the cleaned bushing installed. However, the PAH measuring chamber can be
used if the PAH calibration chamber is missing.
Module Type
Calibration Liquid µg/l
Validation
Liquid µg/l
Low Range
Calibration
High Range Calibration
Low
Range
High
Range
Low
Calibration
High
Calibration
Low
Calibration
Middle
Calibration
High
Calibration
G6110
Low
Range
0
100
-
-
-
50
-
G6111
G6110
High
Range
-
-
0
400
800
-
400
G6111
G6110
Dual
Range
0
100
100
400
800
50
200
G6111
-
Table 6-3: Validation/Calibration Liquid Selection